Respiration Monitoring Condition Setup
●
RR Synchronization Mark
The RR mark synchronized to impedance respiration, multigas unit, or ventilator will be displayed
inside the numeric data box.
1
Select ON or OFF for “RR Sync. Indicator”.
ON will display the synchronized mark.
OFF will not display the synchronized mark.
●
RR Alarm Source
The parameter to display the RR synchronized mark and to generate the RR/Apnea alarm can be
selected from the Impedance RR, Gas RR, and Ventilator RR.
The RR Alarm Source will be displayed inside the respiration numeric data box.
1
Select the RR Alarm Source.
Impedance will generate RR alarm and display the RR
synchronization mark based on the impedance respiration
curve.
GAS will generate RR alarm and display the RR
synchronization mark based on the RR measured by the
multigas unit.
Vent. will generate RR alarm and display the RR
synchronization mark based on the RR measured by the
ventilator.
C A U T I O N
If the RR alarm source is other than the impedance respiration, the respiration
waveform will not be transmitted on the wired network.
RR Sync. Indicator
<Impedance RR>
<Gas RR>
<Ventilator RR>
Содержание DynaScope 7000 Series
Страница 38: ...xxxvi Blank Page ...
Страница 126: ...4 78 Blank Page ...
Страница 148: ...5 22 Blank Page ...
Страница 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Страница 322: ...7 78 Blank Page ...
Страница 428: ...10 32 Blank Page ...
Страница 466: ...11 38 Blank Page ...
Страница 474: ...12 8 Blank Page ...
Страница 475: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4L010175E 201207 ...