Cleaning/Sterilizing the Airway Adapter (IRMA Probe)
The IRMA probe can be cleaned using a cloth moistened with ethanol or isopropyl alcohol.
C A U T I O N
z
The IRMA oxygen sensor cell and IRMA airway adapters are non-sterile
devices. Do not autoclave the devices as this will damage them.
z
Never sterilize or immerse the IRMA probe in liquid.
Replacing the Air Filter
The cooling fan air filter on the DS-7000 and Multigas Unit is a consumable product.
Continuous operation of the fan will cause the air filter to inhale unclean air inside the equipment,
reduce the cooling effect, and may damage the inner parts.
Clean the air filter, or replace it with a new air filter every 3 months.
W A R N I N G
z
After washing the air filter with neutral detergent, dry it completely before
reattaching. If moisture remains on the air filter, it may damage the
equipment.
z
The air filter must be attached after cleaning / replacing. If the equipment
is used without it, the equipment will get damaged.
●
Cleaning and Replacing the Air Filter of the DS-7000 Main Unit
1
Turn the cooling fan cover to the “UNLOCK” direction, and take it off when the
△
mark
matches the arrow mark on the monitor.
10
Clea
ning
2
Take out the air filter and clean it, or replace it if necessary.
To clean the air filter, beat the dust off, or wash it off with neutral detergent.
Air Filter
3
Reattach the cooling fan cover.
Turn to the “LOCK” direction and attach it so that the
△
mark and the arrow mark aligns.
10
-
7
Содержание DynaScope 7000 Series
Страница 38: ...xxxvi Blank Page ...
Страница 126: ...4 78 Blank Page ...
Страница 148: ...5 22 Blank Page ...
Страница 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Страница 322: ...7 78 Blank Page ...
Страница 428: ...10 32 Blank Page ...
Страница 466: ...11 38 Blank Page ...
Страница 474: ...12 8 Blank Page ...
Страница 475: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4L010175E 201207 ...