6
-
38
Pulse Wave Size
1
Select the waveform size.
Select the waveform size for displaying and recording from
×
1/4 ,
×
1/2 ,
×
1 ,
×
2 ,
×
4 .
SpO
2
Alarm
1
Press the SpO
2
Alarm key to display the alarm setup menu.
Select ON/OFF for SpO
2
alarm, and set the upper and lower alarm limit.
Also, when the SpO
2
value is unstable around the lower alarm limit, the frequently generated alarm
can be corrected by setting the SEC (second) alarm function.
The upper and lower limits can be set in 1% increment.
Key
Item
Description
Individual Alarm
Selecting ON will generate the SpO
2
alarm.
Selecting OFF will not generate the SpO
2
alarm.
▲
Current Value
Indicates the current value.
Lower Alarm Limit
Sets the lower alarm limit (70 to 99%).
Setting the value to 70% or below will turn OFF the alarm.
Upper Alarm Limit
Sets the upper alarm limit (71 to 100%).
Setting the value to 100% or above will turn OFF the
alarm.
Automatic Setup
Automatically sets the upper limit to OFF, and the lower
limit to 90%.
N O T E
Whether to use the SEC (second) alarm function and its threshold selection should
be based on the patient’s clinical indication portent and medical evaluation.
To maintain the alarm setting even after the power is turned OFF or after the discharge procedure,
store the setting to one of the alarm modes, or select ON for “Backup at Discharge” on the
Discharge Mode setup menu under the Initial Settings.
Reference
For the alarm mode setup procedure, refer to “9. Initial Settings –Alarm Mode Setup–”.
Содержание DynaScope 7000 Series
Страница 38: ...xxxvi Blank Page ...
Страница 126: ...4 78 Blank Page ...
Страница 148: ...5 22 Blank Page ...
Страница 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Страница 322: ...7 78 Blank Page ...
Страница 428: ...10 32 Blank Page ...
Страница 466: ...11 38 Blank Page ...
Страница 474: ...12 8 Blank Page ...
Страница 475: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4L010175E 201207 ...