Solution : Connect the airway adapter to the IRMA probe.
If the trouble persists, contact our service representative.
The “Zeroing Required” message is displayed.
Cause
: The gas parameters are out of zero reference.
Solution : Perform zero calibration.
Refer to “3. Vital Application To Measure the Multigas (IRMA Probe) Zero Calibration”
If the trouble persists, contact our service representative.
Oximeter
The measurement data is not displayed.
Cause 1 : The cable is not properly connected.
Solution : Securely connect the following cable to the serial connector and the corresponded
device.
Oximeter
Continuous Cardiac Output
Measurement Device
External Output Cable
Vigilance CJO-04RS4
Vigilance CEDV
CJO-04RS4
Vigilance
ΙΙ
CJ-502
Vigileo CJ-502
OXIMETRIX3 CJ-508
Q-vue CJO-04RS4
Q2 Computer
CJO-04RS4
Cause 2 : The serial communication setup is not properly set.
Solution : Set the connected port function to Vigilance , Oximetrix3 , Q-vue or
Q2Computer on the serial communication setup.
Cause 3 : The measurement data is not displayed on the oximeter display.
Solution : The measurement data of SvO
2
, CO, etc. will not be displayed on the monitor unless the
data is displayed on the used oximeter. Check if the data is displayed on the used
oximeter.
Cause 4 : The CCO is not measured.
Solution : The monitor will display CCO/CCI data only during the process of CCO measurement on
the oximeter.
Cause 5 : The BSA is not input.
Solution : To display the CCI data on the monitor, it is necessary to input the BSA to the Q-vue/Q2
Computer. To display the CI AVG data, it is necessary to input the CO AVG and BSA to
the Q-vue/Q2 Computer. For procedures, refer to the operation manual of the Q-vue/Q2
Computer. For OXIMETRIX3, CI AVG, BSA cannot be displayed on the monitor as BSA
cannot be received from the OXIMETRIX3.
Cause 6 : The communication setup of the DS-7000 and the oximeter is not corresponded.
Solution : The communication setup of the DS-7000 is fixed to the default setting for each oximeter
and cannot be changed. Check if the communication setup of the oximeter is in default
setting.
•
In case of Vigilance/Vigileo, check if the communication is set as follows.
・
Device: IFM Out
・
Baud Rate: 19200bps
・
Parity Bit: None
・
Stop Bit: 1
・
Data Bit: 8
・
Flow Control: 2 sec.
For procedure to check the Vigilance/Vigileo communication setup, refer to the operation
manual for the Vigilance/Vigileo.
10
-
30
Содержание DynaScope 7000 Series
Страница 38: ...xxxvi Blank Page ...
Страница 126: ...4 78 Blank Page ...
Страница 148: ...5 22 Blank Page ...
Страница 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Страница 322: ...7 78 Blank Page ...
Страница 428: ...10 32 Blank Page ...
Страница 466: ...11 38 Blank Page ...
Страница 474: ...12 8 Blank Page ...
Страница 475: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4L010175E 201207 ...