C A U T I O N
z
Precautions about the DS-LANIII Network System
y
In order to connect to the DS-LANIII network, the software version
needs to be the version which supports the DS-LANIII. For details, refer
to our service representative.
y
Make sure that “DS-LAN Setup” for all bedside monitors and central
monitors are set to DS-LANII before connecting the monitor to the
network.
y
If the measurement unit for BP (mmHg/kPa) and temperature (
°
C/
°
F) is
different between the bedside monitor and the central monitor, the
corresponding waveform and numeric data will not be displayed on the
central monitor.
y
If using a HUB for the DS-LAN III network construction, make sure to
use a switching HUB recommended by Fukuda Denshi.
y
The displayable waveform, numeric data, and alarm will differ
depending on the central monitor model type. Refer also to the
operation manual for the respective central monitor.
y
On the DS-LANIII system, the patient data search function using the
patient data server can be used depending on the monitor model type
and software version. For details, refer to our service representative.
z
When DS-5800N/NX/NX
MB
is used as a central monitor, recall,
trend/table, data will not be displayed. Also,
Σ
recording cannot be
performed. For the ST display, overlap waveform will not be displayed on
the DS-5800N/NX/NX
MB
until 15 minutes elapses since the reference
waveform is set on the DS-7000.
z
As the DS-7000 do not have the arrhythmia template display and 12-lead
ST display function, these displays on the central monitor will not be
supported.
z
If connected to a wired network, time/date will be the same with the
central monitor. Even if the time/date is changed on the DS-7000, it will be
synchronized with the time/date of the central monitor.
z
If ECG lead (ECG1 or ECG 2) is changed on the DS-7000 while
monitoring ST display on the central monitor, the ST display will be
distorted. Redrawing the ST display will return the display to normal.
z
The respiration waveform and RR value based on the RR source selected
on the DS-7000 will be displayed on the central monitor. The monitoring
RR and APNEA will be the same as the one monitored on the DS-7000.
xxvii
Содержание DynaScope 7000 Series
Страница 38: ...xxxvi Blank Page ...
Страница 126: ...4 78 Blank Page ...
Страница 148: ...5 22 Blank Page ...
Страница 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Страница 322: ...7 78 Blank Page ...
Страница 428: ...10 32 Blank Page ...
Страница 466: ...11 38 Blank Page ...
Страница 474: ...12 8 Blank Page ...
Страница 475: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4L010175E 201207 ...