6
-
45
SpO
2
Monitoring Condition Setup
Pressing the Setup key on the SpO
2
configuration menu will allow setting the SpO
2
measurement
condition.
●
SpO
2
Alarm during NIBP Measurement
This setup can be used when the SpO
2
sensor and the NIBP cuff is placed on the same limb for
measurement.
During the NIBP measurement, the cuff inflation restricts the blood flow which disables the correct
detection of SpO
2
and may generate an improper alarm.
Selecting OFF for “Alarm during NIBP” will not a generate the SpO
2
alarm until the NIBP
measurement is complete. Similarly, when the PR source is SpO
2
,
the PR alarm will not be generated
during NIBP measurement.
1
Select ON or OFF for “Alarm during NIBP”.
ON will generate an alarm during NIBP measurement.
OFF will not generate a SpO
2
/PR alarm during NIBP
measurement.
●
PR Source (BP or SpO
2
)
The measurement source for PR can be selected. The PR synchronized mark will be also displayed
with the selected source.
1
Select the PR source.
SpO
2
/BP will prioritize SpO
2
as measurement source
for PR. When PR cannot be measured with SpO
2
, the
measurement source will switch to BP.
BP/SpO
2
will prioritize BP as measurement source for
PR. When PR cannot be measured with BP, the
measurement source will switch to SpO
2
.
C A U T I O N
If “HR/PR Alarm Source” is PR, and “PR Source” is BP, the ECG waveform
will not be transmitted on a wired network. On the central monitor, the PR-IBP
value will be displayed for HR. However, the HR value from ECG will be
displayed for the NIBP list and ST measurement table.
N O T E
z
If PR source is set to BP, make sure to measure the corresponded BP
(BP1 or ART). Otherwise PR will be displayed as “
−
−
−
”.
z
The “Sync. Mark/Tone” (ECG setup) and “HR/PR Alarm Source” settings
will be synchronized with the HR/PR user key function.
6
SpO
2
(Ma
sim
o
®
Model; DS
-70
00M
)
Содержание DynaScope 7000 Series
Страница 38: ...xxxvi Blank Page ...
Страница 126: ...4 78 Blank Page ...
Страница 148: ...5 22 Blank Page ...
Страница 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Страница 322: ...7 78 Blank Page ...
Страница 428: ...10 32 Blank Page ...
Страница 466: ...11 38 Blank Page ...
Страница 474: ...12 8 Blank Page ...
Страница 475: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4L010175E 201207 ...