6
-
65
Temperature Alarm
1
Press the TEMP
*
Alarm key.
The alarm setup menu will be displayed.
Select ON/OFF for the temperature alarm, and set the upper
and lower alarm limit.
The alarm limit can be set for each measurement unit (
°
C /
°
F).
The upper and lower limit can be set in increments of 0.5
°
C /
1.0
°
F.
Key
Item
Description
Individual Alarm
Selecting ON will generate a TEMP alarm.
Selecting OFF will not generate a TEMP alarm.
▲
Current Value
Indicates the current value.
Lower Alarm Limit
Sets the lower alarm limit (30.0 to 49.0
°
C / 86.0 to
120.0
°
F). Setting the value 30.0
°
C / 86.0
°
F or below will
turn the alarm OFF.
Upper Alarm Limit
Sets the upper alarm limit (31.0 to 50.0
°
C / 88.0 to
122.0
°
F). Setting the value 50.0
°
C / 122.0
°
F or above will
turn the alarm OFF.
Automatic Setup
Automatically sets the upper limit to +2.0
°
C / +4.0
°
F from
the current value, and lower limit to –2.0
°
C / –4.0
°
F from
the current value.
To maintain the alarm setting even after the power is turned OFF or after a discharge procedure, store
the setting to one of the alarm modes, or select ON for “Backup at Discharge” on the Discharge
Mode setup menu under the Initial Settings.
Reference
For the alarm mode setup procedure, refer to “9. Initial Settings –Alarm Mode Setup–”.
∆
T Display
The difference between the TEMP1 and TEMP2 value can be displayed in an absolute value.
1
Select ON or OFF .
ON will display the
∆
T value.
OFF will not display the
∆
T value.
6
T
emp
er
atur
e
Содержание DynaScope 7000 Series
Страница 38: ...xxxvi Blank Page ...
Страница 126: ...4 78 Blank Page ...
Страница 148: ...5 22 Blank Page ...
Страница 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Страница 322: ...7 78 Blank Page ...
Страница 428: ...10 32 Blank Page ...
Страница 466: ...11 38 Blank Page ...
Страница 474: ...12 8 Blank Page ...
Страница 475: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4L010175E 201207 ...