Ventilator
P-V, F-V Loop Display
By connecting a ventilator, P-V loop (airway pressure / volume) and F-V loop (airway flow / volume) can
be monitored.
N O T E
If “Servo-i” is not selected on the serial communication setup menu, “Ventilator” key
will not be displayed.
P-V Loop
The P-V loop is sampled every 60ms and displayed for each respiration. The beginning of the loop
trace is displayed in white, and the rest of the loop is displayed in cyan. The horizontal axis shows AWP
(Unit: cmH
2
O), and the vertical axis shows Volume (Unit: mL).
1
Press the Menu
→
Function
→
Ventilator keys.
The P-V loop display will appear.
2
Select the scale for P-V loop.
The scale will change as the Scale key is pressed.
・
Vertical Axis (Volume)
Select from 250 / 500 / 750 / 1000 (mL).
・
Horizontal Axis (Pressure)
Select from 10 / 20 / 30 / 50 / 120 (cmH
2
O).