HR Average Method
The averaging method of HR measured from ECG can be selected.
1
Select Instant or Average for “HR Average” on the 2nd page of the ECG setup menu.
Instant will display the HR measured from RR interval for
each heartbeat.
Average will display the average HR measured for 6
seconds for adult and child, and 3 seconds for neonate.
●
Noise Detection
1
Set the “Noise Detection” on the 2nd page of the ECG setup menu.
ON will automatically switch the HR numeric data and HR
synchronized tone to PR when electrical surgery noise or
motion artifact is detected.
If PR is not measured, HR data before the noise detection will
remain displayed.
During this condition, HR synchronized tone will not be
generated.
OFF will not switch HR to PR even when noise is detected.
When ON is selected for “Noise Detection”
If some noises are detected and PR is also measured, HR
data will switch to PR.
The measurement data will be surrounded by a red box to
indicate that an artifact is present.
If PR is not measured, then the HR data before the noise
detection remains displayed (with HOLD showing inside the
parameter box).
The measurement data will be surrounded by a red box to
indicate that an artifact is present.
C A U T I O N
Arrhythmia may be erroneously detected as noise.
6
ECG
Содержание DynaScope 7000 Series
Страница 38: ...xxxvi Blank Page ...
Страница 126: ...4 78 Blank Page ...
Страница 148: ...5 22 Blank Page ...
Страница 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Страница 322: ...7 78 Blank Page ...
Страница 428: ...10 32 Blank Page ...
Страница 466: ...11 38 Blank Page ...
Страница 474: ...12 8 Blank Page ...
Страница 475: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4L010175E 201207 ...