The “SpO
2
Unit Error” message is displayed.
Cause
: There is a communication failure with the SpO
2
measurement unit.
Solution : A defective cable or SpO
2
unit failure can be considered.
Contact our service representative.
The “Disconnect” message is displayed.
Cause
: When the SpO
2
relay cable is disconnected during SpO
2
monitoring, this message will be
displayed.
Solution 1 : To silence the alarm, press the
(Alarm Silence) key.
Solution 2 : To continue monitoring, plug in the SpO
2
relay cable. This will clear the message and
silence the alarm.
Non-Invasive Blood Pressure
The cuff does not inflate although the pump is operating.
Cause 1 : The air hose is not firmly connected, and the air is leaking.
Solution : Check if the air hose is properly connected.
Cause 2 : The cuff size does not correspond with the selected patient type.
Solution : Check if the cuff size corresponds with the selected patient type.
The monitor repeats the measurement, or “
ー
ー
ー
” is displayed for the numeric data.
Cause 1 : The measurement accuracy is not reliable due to body motion artifact.
Solution : Have the patient stay still as much as possible during the measurement.
Cause 2 : The pulse is too small to acquire reliable measurement accuracy.
Solution : Check if the cuff application is done properly, and if the cuff size corresponds with the
selected patient type.
10
The time of measurement disappears and the numeric data is displayed as “
ー
ー
ー
”.
Trou
ble
sho
oting
Cause
: The NIBP data will be deleted when the preprogrammed NIBP erase time has elapsed.
Solution : Select the appropriate time for NIBP data erase time from 60min or 120min which best
fits the monitoring purpose.
Temperature
The numeric data is displayed as “
×××
”.
Cause
: The temperature measurement is outside the measurement range.
Solution : Check if the temperature probe is properly attached.
The “TEMP not connected” message is displayed.
Cause
: When the temperature sensor is disconnected during temperature monitoring, this
message will be displayed.
Solution 1 : To silence the alarm, press the
(Alarm Silence) key.
Solution 2 : To continue monitoring, plug in the temperature sensor. This will clear the message and
silence the alarm.
“TEMP Unit Error” message is displayed.
Cause
: A failure in the temperature unit is detected.
Solution : A unit failure can be considered. Cease the measurement and contact our service
representative.
“Noise Interference” message is displayed.
Cause
: The electrosurgical noise interference has been detected for 30 seconds and more.
10
-
19
Содержание DynaScope 7000 Series
Страница 38: ...xxxvi Blank Page ...
Страница 126: ...4 78 Blank Page ...
Страница 148: ...5 22 Blank Page ...
Страница 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Страница 322: ...7 78 Blank Page ...
Страница 428: ...10 32 Blank Page ...
Страница 466: ...11 38 Blank Page ...
Страница 474: ...12 8 Blank Page ...
Страница 475: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4L010175E 201207 ...