86
Splice Mode
Parameter
Description
Gapsetting
Gap Set
[Splicing] [L Only] [R Only] [Manual] changes Gap Set
Cleaning Arc
Power
Note1
Sets the Cleaning Arc Power.
[STD] or [SP] or [NC] is selected automatically in Arc Calibration
Method.
Cleaning Arc
Time
Note1
Adjusts duration of the cleaning arc. The arc power of [Cleaning Arc] is the
power set in the field [Prefuse Power].
Gap
Sets the end-face gap between the left and the right fibers at the time of
aligning and prefusion discharge. When splicing fibers of small outside
diameter, fusing by prefusion discharge can be reduced by diminishing this
gap. Adjusting the prefuse power and prefuse time is usually easier method.
Gapset Position
Sets the relative position of the splicing location to the center of electrodes.
Splice loss may be improved in the case of dissimilar fiber splicing by
shifting [Gapset Position] towards a fiber whose MFD is bigger than the
other fiber MFD.
Note 1: FSM-100M/P only
FSM-100M
+
/P
+
Only
Parameter
Description
Cleaning Arc
Cleaning Arc
Power
Sets the Cleaning Arc Power
[STD] or [SP] or [NC] is selected automatically in Arc Calibration Method.
Cleaning Arc
Time
Adjusts duration of the cleaning arc. The arc power of [Cleaning Arc] is the
power set in the field [Prefuse Power].
Electrode swing
Set ON or OFF. 2 items below are needed when ON is selected.
Swing home
position
This is to set swing home position.
Swing stroke
This is to set electrode swing stroke
Both left and right
fibers move to the gap
set position.
Left fiber moves to the
gap set postion.
Right fiber moves to
the gap set position.
Splice
L Only
R Only