167
Maintenance Menu
12.
Maintenance Info.
Select the [Maintenance Info]. The following information is displayed.
Parameter
Description
Serial Num.
Displays the serial number of the splicer.
Version.
Displays the version number of the software.
Arc Count
Displays the number of arc discharges after electrode
replacement. Performing the function [Replace Electrodes] or
[Arc Count Clear] resets this parameter to zero.
This value is highlighted in yellow when it exceeds the
[Electrode Caution] or in red when it exceeds the [Electrode
Warning].
Total Count
Displays the total number of arc discharges.
Electrodes Time
Used
Displays the time of arc discharge after electrode replacement.
Performing [Replace Electrodes] or [Arc Count Clear] resets
this value to zero.
This value is highlighted in yellow when it exceeds the
[Electrode Caution] or in red when it exceeds the [Electrode
Warning].
Last Authorized
Service
Displays the date of last authorized service.
Next Authorized
Service
Displays the scheduled date of next authorized service.
Содержание FSM-100M
Страница 2: ...FSM 100M P series FSM 100M P series ...