160
Maintenace Menu
6.
Cleaning Electrode (FSM-100M+/P+ Only)
Operation Procedure
(1)
Press MENU, select the [Cleaning Electrode] in [Maintenance Menu].
(2)
Open the wind protector, up/down the electrodes with pressing the UP DOWN
key.
(3)
Clean the electrode tip using the electrode cleaner.
•
This function is applicable to LDF electrode (ELCT3-25LDF, 60 degrees
tip) for FSM-100M+/P+.
• Do not use this function for cleaning the electrode (ELCT3-25). Because
electrode tip will end and arc discharge is not stable when it is cleaned.
Replacement is necessary.
• It is recommended that the electrodes are cleaned after 100 arc discharges.
Make sure if the electrode is inserted fully at the cleaner ring part, then push
the cleaner to the electrode and press “ON” button on the cleaner.
If the electrode is NOT fully inserted, cleaner head may hit v-groove and
lens, and result in damage.
Please carefully check if the electrode is fully inserted.
Cathode electrode