95
Splice Mode
[Special Function]
Parameter
Description
Step [1~10]
Edit Control
Parameter
Press the ENT key to edit Special Function parameters. Parameters
may be edited for Step 1 to Step 10 of a Special Function sequence.
Preceding Action
Action
Select [None], [Config Arc], or [Pause] as the Preceding Action.
[None] [Config Arc]
[None]:
No preceding action is taken. This step proceeds normally.
[Config Arc]:
If selected, the [Electrode Gap] and/or [V Height Shift]
parameters may be used to change the arc configuration from the
previous setup.
[Pause]:
If selected, the splicer pauses before proceeding with this step.
The pause allows the operator to select options for how to proceed.
NOTE:
Details concerning use of pause during Special Function operation is
shown at the end of the Special Function section
Electrode
Gap
Set the [Electrode Gap] from 1 mm up to 3 mm.
This allows changing the [Electrode Gap] from the previous setting.
This parameter appears only if [Config Arc] is selected as the [Action].
V Height
Shift
Set the [V Height Shift] to change the fiber vertical position relative to
the electrodes.
This allows changing fiber vertical position from the previous setting.
This parameter appears only if [Config Arc] is selected as the [Action].
Edit Control Parameter
Special Function: ON
[Electrode Gap] &
[V Height Shift]
parameters appear
only if [Config
Arc] is selected as
Preceding Action