172
Error Message List
Error Message
Reason
Solution
Focus X
Motor
Overrun
(Forward)
•The fiber is not set
correctly at the bottom
of the V-groove. The
fiber position is out of
focus range.
•Dust or dirt is on the
fiber surface resulting in
focus failure.
•Dust or dirt is on the
objective lens or the
wind protector mirror.
•Press RESET key, and re-position the
fiber again to seat it correctly at the
bottom of the V-groove.
• If clamping onto coating and the edge of
coating is in the V-groove, prepare the
fiber again.
If edge of coating is still on the V-groove
after above operation, confirm setting of
Tools.
If possible, use the glass clamping mode
to increase stability.
•Confirm the [Coating Diameter] setting
and the [Clad Diameter] setting.
•Prepare the fiber again.
•Execute the [Dust Check]. Clean the lens
or mirror if dust or dirt exists.
Focus X
Motor
Overrun
(Backward)
Focus Y
Motor
Overrun
(Forward)
Focus Y
Motor
Overrun
(Backward)
X-Motor
Overrun
(Upward)
•The fiber is not set
correctly at the bottom
of the V-groove. The
fiber is offset too far and
it exceeds the X or Y
motor range.
•Press RESET key, and re-position the
fiber again to seat it correctly at the
bottom of the V-groove.
•Confirm the [Coating Diameter] setting
and the [Clad Diameter] setting.
• If clamping onto coating and the edge of
coating is in the V-groove, prepare the
fiber again.
If edge of coating is still on the V-groove
after above operation, confirm setting of
Tools.
If possible, use the glass clamping mode
to increase stability.
X-Motor
Overrun
(Downward)
Y-Motor
Overrun
(Upward)
Y-Motor
Overrun
(Downward)