83
Splice Mode
Parameter
Description
Fundamental Settings
Fiber Type
Compare
[Fiber Type Compare]: Fiber image data is used to ensure that the
correct types of fiber are loaded for a particular splice mode.
[Ignore LR]: The splicer ensures the correct fiber types have been
loaded, but ignores left/right fiber position
[Judge LR]: The splicer confirms correct fiber type and also ensures
that the fibers are set on the correct side of the splicer
[Fiber Data]: Select and link appropriate fiber recognition data from
the [Fiber Type Compare] database.
See section [
Fiber Data L
[Fiber Data Learning]→ [Fiber Type Compare]: Load fiber data.
Fiber Data R
Precise Motor
Drive
This function enables enhanced motor movement precision for certain
manual operations. When this function is ON, if the X/Y motor is
moved, the PAS camera system measures the exact fiber movement to
ensure the movement distance is correct. When [Precise Motor Drive]
is ON, the Precise Motor Drive menu appears just after automatic X/Y
alignment.
Proof Test
Selects Proof Test settings.
Electrode Wear
Coef.
Sets an increasing ratio of [Electrodes Time Used]. The value is
determined by multiplying the arc time and this coefficient and is
added to [Electrodes Time Used] on every arc discharge.
Splice Before
Shaping
If this parameter is set to [Yes], a splice of a left and right fiber will be
performed (according to the programmed splice mode parameters) at
the first step when [Fiber Shaping] has been selected as the [Operating
Mode].
Splice quality will not be inspected, loss estimation will not be
performed and error messages such as “Thin Fiber”, “Fat Fiber” (etc.)
will not appear. Instead, the splicer will proceed to the [Fiber Shaping]
steps as programmed in the [Special Function] parameters.
This capability is useful for various special activities, such as an
automated operation to splice a coreless fiber to a SM fiber, and then
automatically generate a ball lens with the coreless fiber.
NOTE:
This parameter only appears if [Fiber Shaping] has been
selected as the [Operating Mode].
Clamp Action
This parameter determines what action is taken by Va and clamp
motors during [Reset] or when the wind protector is opened.
[Auto]: Fiber clamps and Va motors reset automatically
NOTE:
[Auto] is the default condition
[Fixed]: Fiber clamps and Va motors remain stationary at the preset
positions determined by left/right fiber data (such as fiber cladding and
coating diameter) unless these motors have been moved manually
Содержание FSM-100M