162
Maintenance Menu
7.
Diagnostic Test
The FSM-100 series has a built in diagnostic test feature that allows the operator
to perform a simple one step evaluation of splicer performance covering several
different critical variables. Perform this function in the event of splicer operation
trouble.
Operation Procedure
(1)
Select the [Diagnostic Test] in the [Maintenance Menu] and execute
[Diagnostic Test]. The following checks will be made.
(2)
Upon completion of all checks and adjustments, a list of results is displayed.
If the Dust Check result is not good, clean the objective lenses. In the case
that cleaning cannot eliminate contamination, there is a possibility that the
contamination may have entered the inside of the optical path. Please
contact the authorized distributor for additional instructions.
The Dust Check and Motor Calibration functions exist as independent
instructions in [Maintenance Menu]. It is possible to execute them
independently.
Check Item
Description
LED Check
Measures and adjusts the brightness of the illumination LED.
Dust Check
Checks the optical path for dust or dirt and judges whether this
disrupts the fiber observation.
If communication exists, this function indicates the location.
Motor Check
Check the Motor Limit Sensor.
Arc Calibration
Automatically calibrates the arc power factor and fiber splicing
position.
I/O Port Check
Checks for normal operation of the input or output terminals of
the internal circuit.
Memory Check
Checks the memory of the internal circuit.
• Before the start of the test, remove the fibers from the splicer.
• When the Motor Check is completed, prepare and load the fibers into
the splicer and press ENT key.