97
Splice Mode
[Special Function]
Parameter
Description
Target
Measurement
Method
Select [Loss Est], [P-Meter], [Axis Offset], or [Diameter] for a Target
[Measurement Method] to be utilized at the end of this Special Function
step, or select [OFF] (no measurement at the end of this step).
Selection of a [Measurement Method] enables
use
of
a
[Measurement
Mode]
and
[Upper/Lower Limit] parameter (shown here) to
control how the Special Function sequence will
proceed at the end of this step.
The [Next Action] after this step will be
determined by the result of the measurement.
[Loss Est]
If [OFF] is selected for [Measurement Method],
no measurement will conducted at the end of this
step, and [Measurement Mode] & [Upper/Lower
Limit] parameters are not available.
Instead, the [Next Action] parameter appears
and this will guide the action following
completion of this step (without any
measurement)
While [Measurement Method] is set to "P-Meter", the electrode swing
function is inactive.
Measurement
Mode
Select a [Measurement Mode] to correspond with the selected [Measurement
Method].
For [Loss Est] or [Axis Offset], select either [Core] or [Clad] as the
[Measurement Mode].
For [P-Meter] select [dB] or [dBm] as [Measurement Mode].
(
NOTE:
Power Meter communications by GPIB must be set up properly.)
For [Diameter], select [Average], [Maximum] or [Minimum] as the
[Measurement Mode].
NOTE:
Fiber image is scanned left-to-right to take diameter measurements
to determine [Average], [Maximum] or [Minimum] diameter.)
L:
Lower Limit
Set Lower and Upper Control Limit value for the Target Measurement.
•
For [Axis Offset] and [Diameter]: Control limit values in µm
•
For [Loss Est]: Control limit values in dB
•
For [P-Meter]; Control limit values in dB or dBm
U:
Upper Limit
These parameters are displayed only if
a [Measurement Method] is selected.
[Next Action] selection is displayed if
[Measurement Method] is [OFF]
[OFF]