116
Splice Mode
Power Meter Alignment by the USB port
Set [External Instrument] to [USB] in [Machine Settings]. Connect the PC, the
splicer, and the power meter. The splicer executes power meter alignment by the
USB port. For the detailed information of the communication method of the PC
and the splicer, see the documents "Fsm100_Command
ReferenceManual(*).pdf", "UsbFsm100Server_Manual(*).pdf" in the folder
"\API" on the Instruction CD. For the detailed information of the communication
method of the PC and the power meter, see the power meter’s manual.
XY Align with Power Meter Feedback
Connect the PC, the splicer, light source, and power meter as shown.
The splicer aligns X- and Y-axis to make the power meter value as strong (weak) as
possible MAX (MIN).
θ Align with Power Meter Feedback
Connect the PC, the splicer, light source, power meter, polarizer, and depolarizer as
shown.
Set [θ Align] to [P Max] or [P Min]. See [Checking or Editing Splice Mode]. In case
of [P Max], the splicer aligns θ-axis to make the power meter value as high as
possible, MAX. On the other hand, in case of [P Min], the splicer aligns θ-axis to
make the power meter value as low as possible, MIN.
PC
Light Source
Power Meter
Fiber
Fiber
USB
Cable
FSM-100M/M+/P/P+
GP-IB
Cable, etc
PC
Light Source
Power Meter
Fiber
Fiber
USB
Cable
FSM-100M/M+/P/P+
GP-IB
Cable, etc
Polarizer
Depolarizer
Содержание FSM-100M