161
Maintenace Menu
(4)
When Anode electrode is to be cleaned, fully open the wind protector by pushing
“wind-protector open lever”, and close the fiber clamp manually. Open the wind
protector anode electrode.
See [Cleaning Wind Protector Mirrors] in [Maintenance of Splicing Quality] for
detail how to open the wind protector.
Clean electrode tip when electrode tip is found worn, rounded, or glass particle is
found at the tip, or arc discharge is not stable. Electrode stabilizing after cleaning
is recommended.
LDF electrode is to be replaced every 1000 arc discharges (1000um cladding
dia.)
Electrode may degrade depending on situation, so earlier electrode replacement
is sometimes recommended.
Anode electrode
Electrode is fully inserted
at the ring shown in the
picture.
Correct fitting
Electrode is not fully
inserted at the ring shown
in the picture.
Loose fitting
Rotate Button
Ring Part