Quick Start Guide
2.
Convenient Functions
2-1.
Arc Method and Arc Calibration
The relation between the arc method and the arc calibration is shown below.
Parameter
Arc Calibration
Special Arc
Calibration
No Calibration
Arc Calibration
Method
STD(Standard)
If STD is not selected, it
doesn't operate.
SP(Special)
If SP is not selected, it
doesn't operate.
NC
(No Calibration)
No calibration
function
Electrode Gap
1.0mm
Existing value
Existing value
V Height Shift
0µm
Existing value
Existing value
Clad Diameter
125µm
Existing value
Existing value
Arc Center
Compensation
Installed
Installed
Not Installed
Calibration
Result
In all modes when STD
is selected, the result of
the arc calibration is
updated to the arc
power.
The result of the arc
calibration is updated to
the arc power in
selected mode.
However, it is necessary
to optimize the arc
power according to the
fiber.
-
Other Setting
[Electrode Gap 1.0mm]
[V Height Shift 0µm]
[Clad Diameter 125µm]
When the arc calibration
is executed, each
parameter automatically
sets above mentioned.
Please execute Special
Arc Calibration when
needed to execute the
calibration in other
settings.
[Electrode Gap]
[V Height Shift]
[Clad Diameter]
The arc calibration is
executed without
changing each setting of
the splice mode that has
been selected.
-