71
Arc Calibration
(4)
The calibration result is displayed after completion.
“Test Finished” message
After arc power and splicing position
calibration are successfully completed press
the ENT key to exit.
“Test Again” message
Arc power and splicing position calibration
are completed but further calibration is
strongly recommended, as the difference
between the previous arc calibrations may
be too large. Press ENT key to perform arc
calibration, or ESC key to exit and stop arc
calibration.
•In some cases, multiple iterations of arc calibration are needed until the
calibration process is successfully completed and the “Test Finish”
message is displayed. Arc calibration can be considered almost
completed if multiple iterations are completed without receiving the
message
•A maximum number of tests can be set so that “Test Finish” message is
displayed specific number of arc calibrations are performed. See the
[
] for detail.