119
Splice Mode
[Step 1]: The fiber is tapered from 125um to 50um in diameter for 7.00mm in
length.
Set the parameters for [Step 1] in [Special Function] as table below.
i) Tapering by ZL/ZR motors
ii) Tapering by Sweep/ZR motors
Parameter
Arc
Arc
Arc Power
SP – 45 bit
SP – 45 bit
Start Time
0 ms
0 ms
Finish Time
T
1
= L
1
×(R
2
+
R
+
1)/(3×V
feed
)
= 7000×(0.4
2
+0.4+1)/(3×0.04)
= 91000 ms
T
1
= L
1
×(R
2
+
R
+
1)/(3×V
feed
)
= 7000×(0.4
2
+0.4+1)/(3×0.04)
= 91000 ms
Parameter
Motor1
Motor2
Motor1
Motor2
Motor
ZL
ZR
Sweep
ZR
Direction
[ Reverse ]
[ Forward ]
[ Left ]
[ Reverse ]
Start Time
0 ms
0 ms
0 ms
0 ms
Finish Time
T
1
T
1
T
1
T
1
Initial Speed
V
feed
V
feed
V
feed
V
feed
Acceleration
a
1
= R/10000
= 0.4/10000
= 0.00004000
0.00000000
(
OFF
)
a
1
= R/10000
= 0.4/10000
= 0.00004000
a
1
= R/10000
= 0.4/10000
= 0.00004000
Acceleration
Type
[ Type 2+ ]
[ Type 1 ]
[ Type 2+ ]
[ Type 3+ ]
L
1
: taper section length in [Step 1] = 7.0mm = 7000 um
R: taper ratio = wrist diameter/fiber diameter = 50/125 = 0.4
V
feed
: fiber feeding speed = 0.04 um/ms (default value)
i) Parameters when tapering by ZL/ZR motors
• If you are not able to taper as designed in the above settings, please change
the “arc power” or “fiber feeding speed”.