117
Splice Mode
For [P Max]
Before θ-axis aligning, adjust the polarizer and the depolarizer so that the value of
the power meter is at maximum.
For [P Min]
Before θ-axis aligning, adjust the polarizer and the depolarizer so that the value of
the power meter is at minimum.
For [XY Align] – [P Meter], and [θAlign] – [P Min]
Before θ-axis aligning, adjust the polarizer and the depolarizer so that the value of
the power meter is at minimum. After finishing θ-axis aligning, the splicer will pause.
At this pause, adjust the polarizer and the depolarizer so that the value of the power
meter is at maximum for XY-axis aligning.
[P Min] aligning can align θ-axis more accurately than [P Max] aligning, because the
power level change more pronounced, sensitive, at the minimum level. However, for
[P Min] aligning, the power meter must measure the power level to less than -80dBm.
• See the instruction manual of the power meter for details of the power meter
parameter settings.
•When using HP8153A/8163, set the sampling time of the selected channel to
20ms. The indicated power level on the HP8153A/8163 is sometimes locked.
This problem is caused by an HP8153A/8163 characteristic. Even if the
indicator is locked, the power meter feed back function works correctly. To
prevent the indicator locking, disconnect the unused channel or set the sampling
time of the unused channel to 60min.