Warning and Caution
《
Placing the splicer in the carrying case, before transportation.
》
Caution:
Transport the splicer in the following order. Otherwise, it may cause equipment
damage. Moreover, the guarantee may not be applied in this case.
(1)
Turn off the power to the splicer.
(2)
Remove the electrodes.
(3)
Insert the Cleaning Stick into the Dust Box and
clean the area.
Remove the dust adhering to the Cleaning
Stick.
(4)
Replace the electrodes.
・
Before placing the splicer in the carrying case, please turn off the power using
the ON/OFF button. Detach the power supply cord before placing splicer in
carrying case.
・
If the power is turned off without using the ON/OFF button, the splicer may not
be ready for transportation. Please do not insert the Z-Unit fixture by force as
this may cause equipment damage.
・
Insert the Cleaning Stick into the Dust Box and clean. Take out the dust and debris
which has stuck to the end of the stick
・
Dust can spread to sensitive machine parts during transportation. As a result, the
splicer might not operate smoothly and repairs may be needed.
・
Do not detach the belt & Z-Unit fixture.
Case of FSM-100M+/P+
・
Tighten the Fixing Screws when the
indicator is aligned with
「
10
」
on the
scale. Please confirm that the “Lock
function” has been activated.
・
The splicer might break down during
transportation if the “Lock function”
has not been activated.
Indicator
Fixing Screw
Содержание FSM-100M