98
Splice Mode
[Special Function]
Parameter
Description
Next Action ([Next Action] Determined by Measurement Result)
Value < L
When [Loss Est], [P-Meter], [Axis Offset], or [Diameter] are selected
for the Target [Measurement Method], the measurement result will
determine the next step. There are 3 possible results:
Measurement result is lower than selected [Lower Limit]
Measurement result is between [Lower Limit] and [Upper Limit]
Measurement result is greater than selected [Upper Limit]
For each case, select a desired [Next Action].
Available choices are as follows:
[Finish]: Special Function process is terminated
[Next Step]: Proceed to next step in Special Function process
[Error]: Process is terminated, Error Message appears
[Repeat]: Repeat current step
NOTE:
If [Repeat] is selected, a [Repeat Limit] parameter appears to
allow control of how many times the step cab be repeated.
[Jump]: Jump to another step in the Special Function process
NOTE:
If [Jump] is selected, a [Step Increment] parameter appears. This
enables control of how many steps forward or backward the process will
jump. [Step Increment] can be set to jump forward up to 9 steps or
backwards up to 9 steps.
Value L to U
Value > U
Because [Jump] is selected for
[Value
U], the [Step Increment]
parameter is available to control
the step to which the process will
jump
Because [Repeat] is selected for
[Value < L], the [Repeat Limit]
parameter is available to control
the maximum number of repeats
Содержание FSM-100M