60
Function Introduction
(3)
Next, splice the fibers under various
conditions to acquire the data.
Measured splice loss and estimated
MFD Mismatch data are needed under
conditions as the following.
Conditions:
Main Arc Time +/-4000ms, +/-3000ms,
+/-2000ms, +/-1000ms (8 splicing data)
(4)
Optimize the MFD Mismatch Sensitivity
so that the dotted line in the graph
becomes same tendency to the sampled
data.
(5)
Finally, optimize the value of Minimum
Loss so that almost all of the sampled
data is in the range of the dotted line in
the graph.
+2sec
-4sec
Optimized condition
[MFD Mismatch Sensitivity]
= 0.50
[Minimum Loss] = 0.16