FT-10 Fill Technical Manual, Rev. 2.21, July 2020
Page 29 of 131
5.4 Programming
5.4.1 Serial Port, Printer and Fieldbuses
[0--] Interface Block
You can reach the parameters about serial interface of FT-10 FILL indicator in this section. The data output
modes can be used once except continuous data output.
[00-] RS 232C Serial Port
This sub-block includes the parameters about the 1
st
serial interface of FT-10 FILL.
[000 3 ] Data Format
0 : No data transfer.
1 : Continuous data output
( * )
(Page 72)
2 : Print mode
(Parameter [
040
]
(Page 32)
3 : BSI command set
(Page 74)
4 : Modbus RTU High-Low
(Page 79)
5 : Modbus RTU Low-High
(Page 79)
6 : Fast continuous mode
(Page 73)
( * ) Warning
:
Use for Flintec remote displays interfacing. CR and LF should be enabled.
[001 3] Baud Rate
0 : 1200 Baud
1 : 2400 Baud
2 : 4800 Baud
3 : 9600 Baud
4 : 19200 Baud
5 : 38400 Baud
6 : 57600 Baud
7 : 115200 Baud
[003 00] Address
You can define a device address between 1 and 99 by this parameter. If you enter 0, controller will
operate without an address.
[004 0] Data length and parity
0 : 8 bit, no parity
1 : 7 bit, odd parity
2 : 7 bit, even parity
3 : 8 bit, odd parity
4 : 8 bit, even parity
[005 0] Checksum
You can enable or disable for continuous data format and BSI command set.
0 : No checksum
1 : Checksum enable
[006 1] Carriage return
You can enable or disable for continuous data format.
0 : No CR
1 : CR enables
[007 1] Line feed
You can enable or disable for continuous data format.
0 : No LF
1 : LF enables
[008 0] Response Speed
0 : Modbus RTU Answer is sent immediately after Request is received.
1 : Modbus RTU Answer is delayed 20 msec after Request is received.
This property is very helpful for slow PLC systems