FT-10 Fill Technical Manual, Rev. 2.21, July 2020
Page 106 of 131
01001011
01011111
75
95
Not Used
01100000
96
R/W
Filling mode
selection
(1)
Refer to par.[500],page 70
01100001
97
R/W
Feeding type
(1)
Refer to par.[501].page 70
01100010
98
R/W
Preact correction
factor
(1)
Refer to par.[502],page 70
01100011
99
R/W
Preact correction freq.
(1)
Refer to par.[503],page 70
01100100
100
R/W
Check delay
(1)
Refer to par.[504],page 71
01100101
101
R/W
Zeroing period
(1)
Refer to par.[505],page 71
01100110
102
R/W
Zeroing delay
(1)
Refer to par.[506],page 71
01100111
103
R/W
Indcation at the end of
filling
(1)
Refer to par.[507],page 71
01101000
104
R/W
Feeding check
(1)
Refer to par.[508] page 71
01101001
105
R/W
Filling time
(1)
Refer to par.[509] page 71
01101010
106
R/W
Start Type
(1)
Refer to par.[50A] page 71
01101011
107
R/W
Indication in filling
proc.
(1)
Refer to par.[50B] page 71
Table 14.3
- Expanded Command List
(1) Write this command with writing values to D0~D33 bits then apply New CMD.
Programming steps of frequent used:
Reading a weight value:
1. Check the D43
…D47 bits of ‘TxPDO 1 (T_UL1)’.
2. If there is not any error, read a weight value (gross, net or tare),
Zero Calibration procedure:
1. Check the low byte of Calibration Status. it should be decimal '1' to start adjustment.
2.
Write ‘Adjust Zero Calibration’ command and apply New CMD to start Zero calibration.
3. Check the low byte of Calibration Status. it is decimal '3' during zero calibration process.
4. The low byte of Calibration Status changes to decimal '1' at the end of the Zero calibration.
5. If the low byte of Calibration Status is '9', check the high byte of Calibration Status to understand
the calibration error.
Span Calibration procedure
:
1. Check the low byte of Calibration Status. it should be decimal '1' to start adjustment.
2.
Write ‘Adjust Span Calibration’ command with writing test weight values to Low Dword of RxPDO1
(R_UL1), then apply New CMD to start Span calibration.
3. Check the low byte of Calibration Status. it is decimal '4' during span calibration process.
4. The low byte of Calibration Status changes to decimal '1' at the end of the Span calibration.
5. If the low byte of Calibration Status is '9', check the high byte of Calibration Status to understand
the calibration error.