FT-10 Fill Technical Manual, Rev. 2.21, July 2020
Page 87 of 131
12.
P
ROFINET
(
ONLY
FT-10
FILL
PN)
The Profinet interface operates at 100Mbit, full duplex, as required by Profinet.
The
GSDML
file is available on Internet
Attention:
There are two different GSDML, v2.2 OR v2.3 files depending on Profinet version.
Profinet interface of the weighing instrument can be done via hub switch or serial bus over two Profinet ports.
1. Serial bus connection. You may connect instruments serial to your Profinet bus via two ports.
2. Star connection. If you connect the instrument to your PLC via hub switch, you can use P1 or P2 port
on the instrument. You may change the port, if there is any malfunction on port in usage.
[06-] Profinet (Only FT-10 FILL PN)
This sub-block includes the parameters related with the Profinet interfaces of FT-10 FILL controller.
[060 0] Data Format
0 : Signed 32 bit integer, no decimal point implied
1 : 32 bit float, decimal point implied
There are 4 announcement LEDs on rear panel to indicate the interface status as seen below. The meanings
of these LED’s are;
MS Module Status LED
LED State
Description
Comment
Off
Not power or not initialized
No power or profinet module is in
initialization state
Green
Normal operation
Green, 1 flash
Initialized, diagnostic event(s)
present
Diagnostic is active
Red
Exception error
There is an exception error
Red, 1 flash
Configuration Error
Check GSDML configuration
Red, 2 flashes
IP Address Error
IP address not set
Red, 3 flashes
Device Name Error
Station name not set
Red, 4 flashes
Internal Module Error
Re-energize the instrument.
If seen again, change the board.
In the case of red LED warning, check cabling, configuration, IP address and device name before
reenergizing the instrument after 30 seconds power off.
LINK/Activity LED
LED State
Description
Comment
Off
No Link
No link, no communication present
Green
Link
Ethernet link established, no
communication present
Green, flickering
Activity
Ethernet link established,
communication present
1
NS Network Status LED
2
MS Module Status LED
3
Link/Activity LED (port 1)
4
Link/Activity LED (port 2)
5
P1 Profinet interface (port 1)
6
P2 Profinet interface (port 2)