FT-10 Fill Technical Manual, Rev. 2.21, July 2020
Page 18 of 131
Warning:
Disconnect IndFace1X PC software before starting Modbus-RTU interfacing.
4.3.5 Profibus Connection (only FT-10 FILL PB)
Profibus connection is done as indicated below in Figure 4.5.
Figure 4.5 - FT-10 FILL PB serial interface connections
PROFIBUS Connector pin configuration (DB9F)
Pin
Signal
Description
1
-
-
2
-
-
3
B Line
Positive RxD / TxD, RS-485 level
4
RTS
Request to send
5
GND Bus
Ground (isolated)
6
+5V Bus Output
+5V termination power (isolated)
7
-
-
8
A Line
Negative RxD / TxD, RS-485 level
9
-
-
Housing
Cable Shield
Ground
4.3.6 Profinet Connection (only FT-10 FILL PN)
Profinet connection is done as indicated below in Figure 4.6.
Figure 4.6 - FT-10 FILL PN serial interface connections
PROFINET Connector pin configuration (RJ45)
Pin
Signal
DIR
Description
1
TX+
Out
Differential Ethernet transmit data +
2
TX−
Out
Differential Ethernet transmit data −
3
RX+
In
Differential Ethernet receive data +
6
RX−
In
Differential Ethernet receive data −
4
Not used
Terminated
5
Not used
Terminated
7
Not used
Terminated
8
Not used
Terminated
Shield
Chasis ground