FT-10 Fill Technical Manual, Rev. 2.21, July 2020
Page 75 of 131
Note:
CHK, CR and LF will not be shown in below data format descriptions in this section.
Commands and Responses:
A
Read all weight data
Command:
[ADR][A]
Response:
[ADR][A][STATUS][SIGN][NET W][SIGN][TARE W][SIGN][GROSS W]
Example:
Command: 01A
Response: 01AS+000123.4+000111.1+000234.5
01AD+000123.4+000111.1+000234.5
01AO (ADC out error)
Comments: The response is net, tare and gross weight values or error status.
All weight data is transmitted immediately after receiving the command.
B
Read Gross weight
Command:
[ADR][B]
Response:
[ADR][B][STATUS][SIGN][WEIGHT VALUE]
Example:
Command: 01B
Response: 01BS+000123.4
(gross weight is stable and 123.4)
01BD+000123.4
(gross weight is dynamic and 123.4)
01B−
(under load)
Comments: The response is the gross weight value (stable or dynamic) or error status.
Gross weight data is transmitted immediately after receiving command.
C
Clear the tare memory
Command:
[ADR][C]
Response:
[ADR][C][A] (Cleared and the scale is in gross mode)
Comments: The response status is always Ack in weighing or force mode.
F
Start /stop continuous data output
Command:
[ADR][F][ENABLE/DISABLE]
Response:
[ADR][F][STATUS]
Example:
Command: 01F1
(Enable)
01F0
(Disable)
Response: 01FA
(Command is done successfully)
01FN
(Command could not be executed)
Comments: Indicated weight value sends continuously.
Continuous data format is [ADR][I][STATUS][SIGN][WEIGHT VALUE].
G
Read voltage value of DC power supply
Command:
[ADR][G]
Response:
[ADR][G][STATUS][VOLTAGE VALUE]
Example:
Command: 01G
Response: 01GA234 (Power supply is 23.4 VDC)
01GA150 (Power supply is 15.0 VDC)
01GA090 (Power supply is 9.0 VDC)
Comments: Voltage value is 3 byte and sends with 0.1 V increment.