FT-10 Fill Technical Manual, Rev. 2.21, July 2020
Page 110 of 131
16.
E
THER
CAT
(
ONLY
FT-10
FILL
EC)
EtherCAT interface of the weighing instrument can be done via hub switch or serial bus over two EtherCAT
ports.
1. Serial bus connection of instruments. You may connect instruments serial to your EtherCAT
bus
via two ports.
2. Star connection. If you connect the instrument to your PLC via hub switch, you can use P1 port
on the instrument.
The EtherCAT interface supports 100Mbit, full duplex operation.
ESI
file for two ports EtherCAT
is available under
There are 4 announcement LEDs on the instrument to indicate the interface status as seen below. The
meanings of these LED’s are;
NS RUN LED
LED State
Description
Off
INIT
Green
OPERATIONAL
Green, blinking
PRE-OPERATIONAL
Green, single flash
SAFE-OPERATIONAL
Red
(Fatal Event)
MS ERR LED
LED State
Description
Off
No error (or no power)
Red, blinking
Invalid configuration; State change received from master is not possible
due to invalid register or object settings.
Red, single flash
Unsolicited state change; Slave device application has changed the
EtherCAT state autonomously.
Red, double flash
Application watchdog timeout
Red
Application controller failure
In the case of red LED warning, check cabling, configuration, IP address and device name before
reenergizing the instrument after 30 seconds power off.
LINK/Activity LED
LED State
Description
Off
No link, no activity
Green
Link sensed, no activity
Green, flickering
Link sensed, activity
1
NS RUN LED
2
MS ERR LED
3
Link/Activity (IN port)
4
Link/Activity (OUT port)
5
EtherCAT (IN port)
6
EtherCAT (OUT port)