FT-10 Fill Technical Manual, Rev. 2.21, July 2020
Page 85 of 131
11.
P
ROFIBUS
(
ONLY
FT-10
FILL
PB)
In Profibus DPV1 interface, baud rate is detected automatically. Supported baud rates are 9.6 kbps, 19.2
kbps, 45.45 kbps, 93.75 kbps, 187.5 kbps, 500 kbps, 1.5 Mbps, 3 Mbps, 6 Mbps and 12 Mbps. No ‘baud
rate’ instance exists.
After programming Profibus related parameters of the FT-10 FILL PB controller, you can communicate with
the instrument. The
GSD
file is available under
[05-] Profibus (Only FT-10 FILL PB)
This sub-block includes the parameters related with the Profibus interfaces of FT-10 FILL controller.
[050 0 ] Data Format
0 : Signed 32 bit integer, no decimal point implied
1 : 32 bit float, decimal point implied
[051 000] Rack Address
The Profibus rack address of FT-10 FILL will be entered via keypad between 001 to 126.
There are two LEDs near the Profibus connector which are;
ST Status LED
State
Indication
Comment
Off
Not power or not initialized
No power or profibus module is in
initialization state
Green
Initialized
Flashing Green
Initialized, diagnostic event(s)
present
Diagnostic is active
Red
Exception error
There is an exception error
OP Operation mode LED
State
Indication
Comment
Off
Not online /No power
Check power and cable
Green
On-line, data exchange
-
Flashing Green
On-line, clear
-
Flashing Red
(2 flash)
PROFIBUS configuration error
Check GSD file configuration.
11.1 Data Format
Data format of weight value can be programmable for Floating point (IEEE 754) or Integer. Refer to
parameter [050].
1
OP Operation mode LED
2
ST Status LED
3
Profibus Connector