FT-10 Fill Technical Manual, Rev. 2.21, July 2020
Page 20 of 131
The PC connection cabling will be done via cross cable as shown below. IP address blocks and gateway
address of FT-10 FILL and PC should be the same in cross connection.
Figure 4.10 - Cross PC connection
Warning
:
Connect the shield to the reference ground or shield pin of the power connector.
Warning:
Disconnect IndFace1X PC software before starting Modbus-RTU interfacing.
4.3.8 CANopen Connection (only FT-10 FILL CO)
CANopen connection is done with four wires as indicated below in Figure 4.11. The data line ends must be
equipped with 120 ohm bus terminating resistors.
Figure 4.11 - FT-10 FILL CO serial interface connections
CANopen Connector pin configuration (DB9M)
Pin
Signal
Description
1
-
-
2
CAN_L
-
3
CAN_GND
-
4
-
-
5
CAN_SHIELD
-
6
-
-
7
CAN_H
-
8
-
-
9
-
-
Housing
Cable Shield
-
Figure 4.12 - FT-10 FILL CO serial interface connector
Warning
:
Connect the shield to the reference ground.
Warning:
Disconnect IndFace1X PC software before starting Modbus-RTU interfacing.