BA00059D/06/EN/15.14
71230715

Valid as of version:
V 3.01.XX (device software, HART 5)
V 3.07.XX (device software, HART 7)

Operating Instructions

Proline Promass 83
HART

Coriolis Mass Flow Measuring System

6

Содержание HART Proline Promass 83

Страница 1: ...ions Services BA00059D 06 EN 15 14 71230715 Valid as of version V 3 01 XX device software HART 5 V 3 07 XX device software HART 7 Operating Instructions Proline Promass 83 HART Coriolis Mass Flow Meas...

Страница 2: ...Proline Promass 83 2 Endress Hauser...

Страница 3: ...iguration 69 6 5 Adjustment 79 6 6 Purge and pressure monitoring connections 83 6 7 Data memory HistoROM F CHIP 83 7 Maintenance 84 7 1 Exterior cleaning 84 7 2 Cleaning with pigs Promass H I S P 84 7...

Страница 4: ...understood these Operating Instructions and must follow the instructions they contain The device must be operated by persons authorized and trained by the facility s owner operator Strict compliance...

Страница 5: ...t information and updates to these Operating Instructions 1 4 Return Do not return a measuring device if you are not absolutely certain that all traces of hazardous substances have been removed e g su...

Страница 6: ...ed order code The device type product root and basic specifications mandatory features are always listed Of the optional specifications optional features only the safety and approval related specifica...

Страница 7: ...cd 5 Power supply frequency and power consumption 6 Additional function and software 7 Available inputs outputs 8 Reserved for information on special products 9 Please refer to operating instructions...

Страница 8: ...easuring tubes Materials 9 Max fluid temperature Tm 10 Pressure range of secondary containment 11 Accuracy of density measurement Density cal 12 Additional information 13 Reserved for information on s...

Страница 9: ...evice software currently installed Device SW 9 Installed communication type Communication 10 Information on current communication software Drivers Device Revision and Device Description 11 Date of ins...

Страница 10: ...he measuring system described in these Operating Instructions thus complies with the statutory requirements of the EC Directives Endress Hauser confirms successful testing of the device by affixing to...

Страница 11: ...of foreign matter to the measuring tube during transportation and storage Consequently do not remove these covers or caps until immediately before installation Do not lift measuring devices of nominal...

Страница 12: ...orting Promass X 3 1 3 Storage Note the following points Pack the measuring device in such a way as to protect it reliably against impact for storage and transportation The original packaging provides...

Страница 13: ...ngs which create turbulence valves elbows T pieces etc as long as no cavitation occurs For mechanical reasons and in order to protect the pipe it is advisable to support heavy sensors 3 2 1 Dimensions...

Страница 14: ...normal conditions In the case of liquids with a low boiling point hydrocarbons solvents liquefied gases or in suction lines it is important to ensure that pressure does not drop below the vapor press...

Страница 15: ...s that no gas bubbles or solids deposits can form in the bent measuring tube single tube system A0018978 Special installation instructions for Promass A Caution Risk of measuring pipe fracture if sens...

Страница 16: ...ble ambient temperature range for the transmitter 131 is not exceeded we recommend the following orientations For fluids with very high temperatures we recommend the horizontal orientation with the tr...

Страница 17: ...direction and at a specific slope gravity can be used to achieve complete drainability The sensor must be installed in the correct position with the tube bend facing to the side to ensure full drainab...

Страница 18: ...The underside of the process connection is indicated by a scribed line This line indicates the lowest point of the eccentric process connection Promass I and P with hygienic connections mounting clam...

Страница 19: ...ovided in the separate Technical Information document 143 3 2 5 Heating Some fluids require suitable measures to avoid loss of heat at the sensor Heating can be electric e g with heated elements or by...

Страница 20: ...the sensors 3 2 6 Thermal insulation Some fluids require suitable measures to avoid loss of heat at the sensor A wide range of materials can be used to provide the required thermal insulation a000461...

Страница 21: ...as far as it will go 4 Turn the transmitter housing to the desired position max 2 90 in either direction 5 Lower the housing into position and reengage the bayonet catch 6 Retighten the two securing...

Страница 22: ...tall the device in a shady location Avoid direct sunlight Always install the wall mount housing in such a way that the cable entries are pointing down Mounted directly on the wall 1 Drill the holes as...

Страница 23: ...lidly seated on the panel wall Afterwards tighten the locking nuts Additional support is not necessary a0001131 Fig 18 Panel installation wall mount housing Pipe mounting The assembly should be perfor...

Страница 24: ...rm the following checks after installing the measuring device in the pipe 4 x 45 Device condition and specifications Notes Is the device damaged visual inspection Does the device correspond to specifi...

Страница 25: ...he housing before the power supply is applied You may only connect the sensor to the transmitter with the same serial number Communication errors can occur if this is not observed when connecting the...

Страница 26: ...isolated power supply Compare the specifications on the nameplate with the local supply voltage and frequency The national regulations governing the installation of electrical equipment also apply 1 U...

Страница 27: ...interface FXA193 FieldCheck FieldCare f Cover of the connection compartment a0001135 Fig 24 Connecting the transmitter wall mount housing cable cross section max 2 5 mm2 a Cable for power supply 85 to...

Страница 28: ...nication boards C Relay output 2 Relay output 1 Frequency output Current output HART D Status input Relay output Frequency output Current output HART E Status input Relay output Current output 2 Curre...

Страница 29: ...T handheld communicator See also the documentation issued by the HART Communication Foundation and in particular HCF LIT 20 HART a technical summary a0004586 Fig 25 Electrical connection of HART handh...

Страница 30: ...le entries and insert plugs instead Do not remove the grommet from the cable entry Caution Do not loosen the screws of the sensor housing as otherwise the degree of protection guaranteed by Endress Ha...

Страница 31: ...ine liquid crystal display shows measured values dialog texts fault messages and notice messages HOME position operating mode is the term given to the display during normal operation Readings displaye...

Страница 32: ...d status variables e g bargraph display 4 Info icons field icons representing additional information on the measured values are shown in this field 33 5 Measured values field the current measured valu...

Страница 33: ...a0001184 Counting mode totalizer BALANCE forward reverse flow a0001185 Counting mode totalizer forward a0001186 Counting mode totalizer reverse a0001188 Volume flow a0001189 Target volume flow a000119...

Страница 34: ...LD or GO ON now appear on the display You can use these to interrupt continue or stop the batching process at any time 30 STOP S Stops batching process HOLD O Interrupts batching process sotkey change...

Страница 35: ...RRENT OUTPUT 1 4 Select a function group e g SETTINGS 5 Select a function e g TIME CONSTANT Change parameter enter numerical values P Select or enter enable code parameters numerical values F Save you...

Страница 36: ...asured values are output via the signal outputs in the normal way If the supply voltage fails all preset and parameterized values remain safely stored in the EEPROM 5 2 2 Enabling the programming mode...

Страница 37: ...and process errors differently by defining them as Fault messages or Notice messages You can define messages in this way with the aid of the function matrix see the Description of Device Functions man...

Страница 38: ...lled commands There are three different command groups There are three different command groups Universal Commands Universal commands are supported and used by all HART devices These are associated wi...

Страница 39: ...r code Power Supply Display option A B C D E F G H X 7 8 HART 5 Valid for software 3 01 XX Function DEVICE SOFTWARE 8100 Device data HART Manufacturer ID Device ID 11 hex ENDRESS HAUSER 51hex Function...

Страница 40: ...Mass flow 27 BALLING 5 Volume flow 28 BRIX 6 Corrected volume flow 29 Other 7 Density 52 Batch up 8 Reference density 53 Batch down 9 Temperature 58 Mass flow deviation 12 Target mass flow 59 Density...

Страница 41: ...Byte 1 Manufacturer ID 17 E H Byte 2 Device type ID e g 66 Promag 53 Byte 3 Number of preambles Byte 4 Universal commands rev no Byte 5 Device specific rev no Commands Byte 6 Software revision Byte 7...

Страница 42: ...to process variables using Command 51 Manufacturer specific units are represented using the HART unit code 240 6 HART 5 Set HART shortform address Access type write Byte 0 desired address 0 to 15 Fac...

Страница 43: ...ange of the primary process variable Bytes 4 7 Upper sensor limit Bytes 8 11 Lower sensor limit Bytes 12 15 Minimum span Note The data relate to the primary process variable Mass flow Manufacturer spe...

Страница 44: ...assigned to long HART tag name Access type read none This command can either be issued with the long frame address of the device or the broadcast address No response if the long HART tag name and the...

Страница 45: ...lue for 4 mA Note Manufacturer specific units are represented using the HART unit code 240 38 Device status reset Configuration changed Access write none none 40 Simulate output current of primary pro...

Страница 46: ...cess variable Byte 1 Device variable code to the second process variable Byte 2 Device variable code to the third process variable Byte 3 Device variable code to the fourth process variable 53 Write d...

Страница 47: ...r of the two indices of the status map array Note Modified mapping is active immediately as soon as this command is executed and can therefore cause immediate changes to the condensed status This comm...

Страница 48: ...tion status active bit is reset the current values that apply for the device status and command 48 response are active Normal internal updating of the device status and command 48 response are resumed...

Страница 49: ...121 I O board and the amplifier board software versions are not compatible 3 6 205 T DAT Data download not successful 3 7 206 T DAT Data upload not successful 4 3 251 Internal communication fault on t...

Страница 50: ...efined batch quantity point was exceeded End of filling process approaching 12 1 474 Maximum flow value entered is overshot 12 7 501 New amplifier software version is loaded Currently no other command...

Страница 51: ...mum current value for the measuring tube exciter coils has been reached since certain process fluid characteristics are extreme 19 2 702 Frequency control is not stable due to inhomogeneous fluid 19 3...

Страница 52: ...6 1 7 357 2 0 358 2 1 359 Pulse output Pulse output frequency is out of range 2 2 360 2 3 361 2 4 362 2 5 379 The measuring tube oscillation frequency is outside the permitted range 2 6 380 2 7 381 Th...

Страница 53: ...error outputs active 15 7 692 Simulation of volume flow active 16 0 700 The process fluid density is outside the upper or lower limit values set in the EPD function 16 1 701 The maximum current value...

Страница 54: ...339 Flow buffer The temporarily buffered flow portions measuring mode for pulsating flow could not be cleared or output within 60 seconds 18 3 340 18 4 341 18 5 342 18 6 343 Frequency buffer The tempo...

Страница 55: ...witched off before you remove the cover of the electronics compartment 1 Switch off power supply 2 Remove the I O board 98 3 Switch HART write protection on or off as applicable by means of the jumper...

Страница 56: ...al and can be switched on via the supply voltage The device then performs internal test functions and the following messages are shown on the local display Normal measuring mode commences as soon as s...

Страница 57: ...Configure another system unit Current output n Freq Pulse output n Selection output type Automatic configuration of display Carrying out another Quick Setup Quit Assign Current output Current span Va...

Страница 58: ...rected volume is derived from the corresponding flow unit 3 The YES option remains visible until all the units have been configured NO is the only option displayed when no further units are available...

Страница 59: ...Setup menu flow fluctuations of this nature can be compensated over the entire flow range and pulsating fluid flows measured correctly You will find detailed instructions on how to use this Quick Setu...

Страница 60: ...4005 B 2002 1003 Esc E XXX XXX XX 1 2 3 4 Yes Yes No No Configure another totalizer Current output n Freq Pulse output n Selection of output type Quit Totalizer 2 Totalizer 1 Totalizer 3 Selection to...

Страница 61: ...G MODE PULSATING FLOW 4208 TIME CONSTANT 0 s Signal type for FREQ PULSE OUTPUT 1 to n for PULSE operating mode 4225 MEASURING MODE PULSATING FLOW Other settings 8005 ALARM DELAY 0 s 6400 ASSIGN LOW FL...

Страница 62: ...sating Flow Quick Setup Note Before carrying out the Quick Setup Batching the Quick Setup Commissioning has to be executed 57 This function is only available when the additional batching software is i...

Страница 63: ...E E ON Value Low flow cut off Flow damping Pressure shock suppression Batch Selector Batch Name Batch Quantity Fix Compensation Quantity Assign Relay Terminal No Max Batch Time QS Batching Dosing YES...

Страница 64: ...ds 6404 PRESSURE SHOCK SUPPRESSION 0 seconds 7200 BATCH SELECTOR BATCH 1 7201 BATCH NAME BATCH 1 7202 ASSIGN BATCH VARIABLE MASS 7203 BATCH QUANTITY 0 7204 FIXED CORRECTION QUANTITY 0 7205 CORRECTION...

Страница 65: ...ft s SI units kg h US units lb min 1 1 24 0 08 0 003 2 1 12 0 4 0 015 4 1 8 1 8 0 066 8 3 8 8 0 3 15 1 2 26 1 0 15 FB 1 2 72 2 6 25 1 72 2 6 25 FB 1 180 6 6 40 1 180 6 6 40 FB 1 300 11 50 2 300 11 50...

Страница 66: ...ids If corrected volume flow e g in Nm3 h is to be measured and output instead of the mass flow e g in kg h change the setting for the CORRECTED VOLUME CALCULATION function to FIXED REFERENCE DENSITY...

Страница 67: ...OW CUTOFF On account of the low mass flow involved when gas flows are measured it is advisable not use a low flow cut off Setting OFF 6402 ON VALUE LOW FLOW CUTOFF If the ASSIGNMENT LOW FLOW CUTOFF fu...

Страница 68: ...measuring points Note For information on installing and removing the T DAT Page 93 a0001221 en Fig 39 Data backup transmission with T DAT SAVE LOAD function Information on the LOAD and SAVE options a...

Страница 69: ...he jumpers 40 Caution Risk of destroying the measuring device Set the jumpers exactly as shown in the diagram Incorrectly set jumpers can cause overcurrents that would destroy either the measuring dev...

Страница 70: ...ing device Set the jumpers exactly as shown in the diagram Incorrectly set jumpers can cause overcurrents that would destroy either the measuring device or external devices connected to it Note that t...

Страница 71: ...ange the setting you must always change the positions of both jumpers Note precisely the specified positions of the jumpers Note that the position of the relay submodule on the I O board can vary depe...

Страница 72: ...et fluid The basic equations without temperature compensation are a0004610 en a0004619 en D1 density of carrier fluid transporting liquid e g water D2 density of target fluid material transported e g...

Страница 73: ...square Expansion coefficient linear Expansion coefficient square Reference temperature Esc E XXX XXX XX Brixgrade density of hydrous saccharose solution in kg m Brix 10 C 20 C 30 C 40 C 50 C 60 C 70 C...

Страница 74: ...e from function COEFFICIENT A0 7032 A1 Value from function COEFFICIENT A1 7033 A2 Value from function COEFFICIENT A2 7034 A3 Value from function COEFFICIENT A3 7035 A4 Value from function COEFFICIENT...

Страница 75: ...6 or 10 9 Assume Density r 870 kg m3 0 870 kg l Temperature T 20 C Coefficients determined for table above A0 2 6057 A1 11 642 A2 8 8571 A3 0 A4 0 B1 2 7747 10 3 B2 7 3469 10 6 B3 0 Calculation a00046...

Страница 76: ...A2 Value from function COEFFICIENT A2 7034 A3 Value from function COEFFICIENT A3 7035 A4 Value from function COEFFICIENT A4 7036 Note Promass determines the reference density by means of the density a...

Страница 77: ...system always by means of the device function REFERENCE CONDITION USER 7401 Caution It is not possible to analyze the trend of process device parameters without reference values Reference values can...

Страница 78: ...amples and pointers for each of the six diagnosis parameters recorded Diagnosis parameters Possible reasons for deviation Mass flow A deviation from the reference status indicates possible zero point...

Страница 79: ...t adjustment can be performed only with fluids that contain no gas or solid contents Zero point adjustment is performed with the measuring tubes completely filled and at zero flow v 0 m s This can be...

Страница 80: ...he code if the function matrix is still disabled CODE ENTRY P Enter the code 83 default CODE ENTRY 83 F Confirm the code as entered The ZERO ADJUST function reappears on the display PROGRAMMING ENABLE...

Страница 81: ...the user has detailed knowledge of the fluid density obtained for example from detailed laboratory analyses The target density value specified in this way must not deviate from the measured fluid dens...

Страница 82: ...NT RUNNING appears on the display for approximately 10 seconds During this time Promass measures the current density of the second fluid measured density value 6487 DENSITY ADJUSTMENT Use P to select...

Страница 83: ...pressure 5 bar 72 5 psi 6 7 Data memory HistoROM F CHIP At Endress Hauser the term HistoROM refers to various types of data storage modules on which process and measuring device data is stored By plug...

Страница 84: ...ial to take the inside diameters of measuring tube and process connection into account see Technical Information 143 7 3 Replacing seals Under normal circumstances fluid wetted seals of the Promass A...

Страница 85: ...r wall mount housing remote version Suitable for Wall mounting Pipe mounting Installation in control panel Mounting set for aluminum field housing Suitable for pipe mounting 3 4 to 3 DK8WM Post mounti...

Страница 86: ...ife cycle management for your plant W M supports you with a wide range of software applications over the entire process from planning and procurement to the installation commissioning and operation of...

Страница 87: ...us signals on display only HART 7 The status signals provide information on the state and reliability of the device by categorizing the cause of the diagnostic information diagnostic event The status...

Страница 88: ...r has occurred are displayed in the TROUBLESHOOTING function Press Enter to acknowledge the errors in question default values are automatically inserted instead of the errored parameter values Note Th...

Страница 89: ...Internal communication fault on the amplifier board Remove the amplifier board 261 S COMMUNICATION I O 261 F No data reception between amplifier and I O board or faulty internal data transfer Check t...

Страница 90: ...o 380 F The measuring tube oscillation frequency is outside the permitted range Causes Measuring tube damaged Sensor defective or damaged Contact your E H service organization 381 S FLUIDTEMP MIN 381...

Страница 91: ...eneous gas or solid content Change or improve process conditions 587 S TUBE OSC NOT 587 F 1 Extreme process conditions exist The measuring system can therefore not be started Change or improve process...

Страница 92: ...esponding diagnosis function 802 S REF D DEV LIM 802 S 1 Advanced Diagnostics The reference density is outside the limit value set in the corresponding diagnosis function 803 S TEMP DEV LIMIT 803 S 1...

Страница 93: ...ulse then restarts the batching Batching via operating keys soft keys The error message is reset by pressing the START key Pressing the START key a second time starts the batching process Batching via...

Страница 94: ...to increase system pressure 1 Install the instrument at the outlet side of a pump 2 Install the instrument at the lowest point of an ascending pipeline 3 Install a flow restriction e g reducer or orif...

Страница 95: ...marily to backflow in the piping because the pulse output cannot subtract in the STANDARD or SYMMETRY measuring modes The problem can be solved as follows Allow for flow in both directions Set the MEA...

Страница 96: ...HOLD VALUE Measured value display on the basis of the last saved value preceding occurrence of the fault ACTUAL VALUE Measured value display on the basis of the current flow measurement The fault is i...

Страница 97: ...serial number printed on the transmitter s nameplate 7 Spare parts are shipped as sets comprising the following parts Spare part Additional parts small items threaded fasteners etc Mounting instructi...

Страница 98: ...de and remove the display module Disconnect the ribbon cable 1 2 of the display module from the amplifier board 3 Remove the screws and remove the cover 2 from the electronics compartment 4 Remove pow...

Страница 99: ...ier board 5 1 Signal cable sensor 5 2 Excitation current cable sensor 5 3 S DAT sensor data memory 5 4 T DAT transmitter data memory 6 I O board flexible assignment 6 1 F Chip function chip for option...

Страница 100: ...nics module 2 Then push up electronics module and pull it as far as possible out of the wall mount housing 3 Disconnect the following cable plugs from amplifier board 7 Sensor signal cable plug 7 1 in...

Страница 101: ...7 Amplifier board 7 1 Signal cable sensor 7 2 Excitation current cable sensor 7 3 S DAT sensor data memory 7 4 T DAT transmitter data memory 8 I O board flexible assignment 8 1 F Chip function chip fo...

Страница 102: ...e fuse is as follows 1 Switch off power supply 2 Remove the power unit board 98 100 3 Remove the protection cap 1 and replace the device fuse 2 Only use the following fuse type 20 to 55 V AC 16 to 62...

Страница 103: ...zardous to health e g flammable toxic caustic carcinogenic etc Note You will find a preprinted Declaration of contamination form at the back of these Operating Instructions 9 8 Disposal Observe the re...

Страница 104: ...surement with temperature compensation Acquisition start via status input for advanced diagnostics SIL 2 New functionalities Operation hours counter Intensity of background illumination adjustable Sim...

Страница 105: ...ess Hauser 105 11 2000 Amplifier 1 00 XX Communication module 1 01 XX Original software Compatible with Fieldtool HART Communicator DXR 275 as of OS 4 6 with Rev 1 DD 1 50098469 11 00 Date Software ve...

Страница 106: ...asured with temperature sensors Measuring range Measuring ranges for liquids DN Range for full scale values liquids gmin F to gmax F mm inch kg h lb min 1 1 24 0 to 20 0 to 0 735 2 1 12 0 to 100 0 to...

Страница 107: ...ax F Max full scale value for liquid kg h lb min r G Gas density in kg m3 lb ft3 for process conditions Here gmax G can never be greater than gmax F Measuring ranges for gases Promass F O Measuring ra...

Страница 108: ...DN 50 Max possible full scale value gmax G gmax F r G x kg m 70000 kg h 60 3 kg m 90 kg m 46900 kg h Recommended full scale values See Page 127 Limiting ow Operable flow range Greater than 1000 1 Flow...

Страница 109: ...lvanically isolated Active 24 V DC 25 mA max 250 mA during 20 ms RL 100 W passive open collector 30 V DC 250 mA Frequency output full scale frequency 2 to 10000 Hz fmax 12500 Hz on off ratio 1 1 pulse...

Страница 110: ...and T DAT save measuring system data if power supply fails HistoROM S DAT exchangeable data storage chip which stores the data of the sensor nominal diameter serial number calibration factor zero poin...

Страница 111: ...w and volume flow liquids 0 10 o r Mass flow gases 0 50 o r Density liquids Reference conditions 0 0005 g cm Field density calibration 0 0005 g cm valid after field density calibration under process c...

Страница 112: ...ressure A difference in pressure between the calibration pressure and the process pressure does not have any effect on the accuracy Design fundamentals Dependent on the flow Flow Zero point stability...

Страница 113: ...Reference conditions 0 0005 g cm Field density calibration 0 0005 g cm valid after field density calibration under process conditions Standard density calibration 0 02 g cm valid over the entire temp...

Страница 114: ...ocess temperature the typical measured error of the sensor is 0 0002 of the full scale value C 0 0001 of the full scale value F Influence of medium pressure The table below shows the effect on accurac...

Страница 115: ...flow and volume flow liquids 0 05 o r PremiumCal for mass flow 0 10 o r Mass flow gases 0 35 o r Density liquids Reference conditions 0 0005 g cm Field density calibration 0 0005 g cm valid after fie...

Страница 116: ...25 o r PremiumCal for mass flow 0 05 o r Mass flow gases 0 25 o r Density liquids 0 00025 g cm3 Temperature 0 25 C 0 0025 T C 0 5 F 0 0015 T 32 F DN Zero point stability Promass F high temperature ver...

Страница 117: ...ndamentals Dependent on the flow Flow Zero point stability Base accuracy 100 Max measured error Base accuracy in o r Repeatability Base accuracy in o r Flow Zero point stability Base accuracy 100 Max...

Страница 118: ...um 2 5W Reference conditions 0 0005 g cm Field density calibration 0 0005 g cm valid after field density calibration under process conditions Standard density calibration 0 02 g cm valid over the enti...

Страница 119: ...is a difference between the temperature for zero point adjustment and the process temperature the typical measured error of the sensor is 0 0002 of the full scale value C 0 0001 of the full scale valu...

Страница 120: ...22 Mass flow and volume flow liquids 0 10 o r Mass flow gases 0 50 o r Density liquids Reference conditions 0 0005 g cm Field density calibration 0 0005 g cm valid after field density calibration unde...

Страница 121: ...0 25 C 0 0025 T C 0 5 F 0 0015 T 32 F Influence of medium temperature When there is a difference between the temperature for zero point adjustment and the process temperature the typical measured err...

Страница 122: ...se frequency output The additional measured error at the current output is typically 5 A Design fundamentals 124 Mass flow and volume flow liquids 0 05 o r PremiumCal for mass flow 0 10 o r Mass flow...

Страница 123: ...or mass flow 0 05 o r Mass flow gases 0 25 o r Density liquids 0 00025 g cm3 Temperature 0 25 C 0 0025 T C 0 5 F 0 0015 T 32 F DN Zero point stability Promass F Standard mm inch kg h or l h lb min 80...

Страница 124: ...e to a difference between calibration pressure and process pressure Design fundamentals Dependent on the flow Flow Zero point stability Base accuracy 100 Max measured error Base accuracy in o r Repeat...

Страница 125: ...density calibration 0 0005 g cm valid after field density calibration under process conditions Standard density calibration 0 01 g cm valid over the entire temperature range and density range 132 Spe...

Страница 126: ...ct on accuracy of mass flow due to a difference between calibration pressure and process pressure Design fundamentals Dependent on the flow Flow Zero point stability Base accuracy 100 Max measured err...

Страница 127: ...density calibration 0 0005 g cm valid after field density calibration under process conditions Standard density calibration 0 01 g cm valid over the entire temperature range and density range 132 Spe...

Страница 128: ...ct on accuracy of mass flow due to a difference between calibration pressure and process pressure Design fundamentals Dependent on the flow Flow Zero point stability Base accuracy 100 Max measured err...

Страница 129: ...conditions 0 0005 g cm Field density calibration 0 0005 g cm valid after field density calibration under process conditions Standard density calibration 0 01 g cm valid over the entire temperature ran...

Страница 130: ...e table below shows the effect on accuracy of mass flow due to a difference between calibration pressure and process pressure Design fundamentals Dependent on the flow Flow Zero point stability Base a...

Страница 131: ...0 C 40 to 140 F Note Install the device at a shady location Avoid direct sunlight particularly in warm climatic regions At ambient temperatures below 20 C 4 F the readability of the display may be imp...

Страница 132: ...5W 50 to 150 C 58 to 302 F Promass I S 50 to 150 C 58 to 302 F Promass E 40 to 140 C 40 to 284 F Promass O 40 to 200 C 40 to 392 F Promass X 50 to 180 C 40 to 356 F Seals Promass F E H I S P O X No i...

Страница 133: ...ersion DN 8 to 25 3 8 to 1 25 bar 362 psi DN 40 to 50 1 to 2 16 bar 232 psi Promass I 40 bar 580 psi Promass P DN 8 to 25 3 8 to 1 25 bar 362 psi DN 40 1 16 bar 232 psi DN 50 2 10 bar 145 psi Promass...

Страница 134: ...value is approx 1 20 of the max full scale value In most applications 20 to 50 of the maximum full scale value can be considered ideal Select a lower full scale value for abrasive substances such as...

Страница 135: ...ight data in lb Promass F DN 8 15 25 40 50 80 100 150 250 Compact version 11 12 14 19 30 55 96 154 400 Compact version high temperature 14 7 30 7 55 7 Remote version 9 10 12 17 28 53 94 152 398 Remote...

Страница 136: ...2 3 4 6 10 Compact version 24 26 31 42 66 121 212 340 882 Compact version high temperature 32 68 123 Remote version 20 22 26 37 62 117 207 335 878 Remote version high temperature 30 65 120 With 10 ac...

Страница 137: ...7 304L Promass E A H I S P Acid and alkali resistant outer surface Stainless steel 1 4301 304 Promass X O Acid and alkali resistant outer surface Stainless steel 1 4404 316L Connection housing sensor...

Страница 138: ...N 11864 1 Form A ISO 2853 SMS 1145 stainless steel 1 4404 316L Tri Clamp OD tubes stainless steel 1 4404 316L Promass A Mounting set for flanges EN 1092 1 DIN 2501 ASME B16 5 JIS B2220 stainless steel...

Страница 139: ...ction DIN 11864 3 Form A stainless steel 1 4435 316L Clamp pipe connection DIN 32676 ISO 2852 BioConnect stainless steel 1 4435 316L Promass O Flanges according to EN 1092 1 DIN 2501 according to ASME...

Страница 140: ...duplex Promass X Stainless steel 1 4404 316 316L manifold 1 4404 316 316L Seals Promass F E H I S P O X Welded process connections without internal seals Promass A Welded process connections without i...

Страница 141: ...approval Information about currently available Ex versions ATEX FM CSA IECEx NEPSI can be supplied by your Endress Hauser Sales Center on request All explosion protection data are given in a separate...

Страница 142: ...ctured according to good engineering practice They correspond to the requirements of Art 3 Section 3 of the Pressure Equipment Directive 97 23 EC Their application is illustrated in Diagrams 6 to 9 in...

Страница 143: ...Supplementary documentation on Ex ratings ATEX FM CSA IECEx NEPSI Special Documentation Functional safety manual Promass 80 83 SD00077D Technical Information Promass 80A 83A TI00054D Promass 80E 83E T...

Страница 144: ...on 6 Device functions See Description of Device Functions manual Display Local display 31 Turning the display 24 Disposal 103 E Electrical connection Cable specifications remote version 26 Commubox FX...

Страница 145: ...tion matrix 35 HART handheld terminal 38 Operational safety 5 Order code Accessories 85 Ordering information 142 Outlet runs 20 Output signal 109 P Performance characteristic Promass A 111 Promass E 1...