Patient Monitor User Manual Monitoring ECG
- 98 -
to PR as the alarm source if:
–a valid ECG lead can no longer be measured and
–a PR source is switched on and available.
The monitor then uses the pulse rate from the measurement currently active as system pulse.
While PR is the alarm source, all arrhythmia and ECG HR alarms are switched off. If an ECG
lead becomes available again, the monitor automatically uses HR as alarm source.
9.6.2 Setting Beat Source
To change the beat source, select either
ECG Setup
>
Beat Source
or
PR Setup
>
Beat Source
.
Select from the following options:
HR
: HR is HR/PR beat source;
PR
: PR is HR/PR beat source;
AUTO
: If the Beat Source is set to
AUTO
, the monitor will use HR as the beat source whenever
the ECG measurement is switched on, and at least one ECG lead can be measured. The monitor
will automatically switch to PR as the beat source if:
•
a valid ECG lead can no longer be measured and
•
a PR source is switched on and available.
If an ECG lead becomes available again, the monitor automatically uses HR as beat source and
the monitor gives a “Di” tone with a blinking heart
displaying in the HR parameter box
when one heartbeat is detected. While a pulse is detected, the monitor gives a “Da” tone.
9.6.3 Smart Lead Off
When
Electrode Type
is
5 Electrodes
,
6 Electrodes
or
10
Electrodes
and
Smart LeadOff
is set
to
On
, if the current selected calculation lead cannot detect ECG signal, the monitor
automatically switches the corresponding lead as calculation lead, and switches the waveform
display of calculation lead at the same time. When ECG electrode is re-connected, and the
original calculation lead recovers its signals, the monitor automatically switch to the original
calculation lead.
To change the smart lead off setting, select
ECG Setup
>
Smart LeadOff
, and select the desired
setting.
NOTE:
When
Electrode Type
is set to
AUTO
, the smart lead off
function will be automatically
turned on and cannot be turned off.
9.6.4 ECG Screen Layout
It varies with
Electrode Type
. When
Electrode Type
is set to
3 Electrodes
,
Screen Layout
can
be set to
Normal
, and it can display one ECG waveform on the main screen.