Patient Monitor User Manual Alarm Information
- 59 -
Message
Cause
Alarm Level
Action Taken
NIBP Excessive
Pressure
Pressure has exceeded the
specified upper safety limit.
Low
Measure again, if
failure persists, stop
measuring function
of NIBP module and
notify biomedical
engineer or
manufacturer’s
service staff.
NIBP Init Pressure
High
The initial pressure is too
high during measuring
Low
NIBP Aux Excessive
Pressure
Pressure has exceeded the
second safety limit as
specified.
High
Notify biomedical
engineer or
manufacturer’s
service staff.
NIBP Time Out
Measuring time has
exceeded the specified time.
Low
Measure again or use
other measuring
method.
NIBP Self Test Error
Sensor or other hardware
errors.
High
Measure again, if
failure persists, stop
using measuring
function of NIBP
module and notify
biomedical engineer
or manufacturer’s
service staff.
NIBP Cuff Type
Error
The cuff type used isn’t
consistent with the patient
type.
Low
Confirm the patient
type and change the
cuff.
NIBP Airway
Pressure
Abnormality
Atmospheric pressure or
system pressure is
abnormal. The valve is
occluded so that deflation is
failed.
Low
Check whether the
airway is occluded
or pressure sensor
works properly. If
the problem still
exists, contact your
service personnel.
Содержание iM50
Страница 1: ......
Страница 29: ...Patient Monitor User Manual Basic Operation 17 iM60 iM70 1 2 3 4 5 6 7 8 9 10 11 1 2 3 11 4 5 6 7 8 9 10...
Страница 317: ......