Patient Monitor User Manual Installation
- 13 -
Chapter 2 Installation
NOTE:
The monitor settings must be configured by the authorized hospital personnel.
2.1 Initial Inspection
Before unpacking, check the packaging and ensure that there are no signs of mishandling or
damage. If the shipping cartons are damaged, contact the carrier for compensation and package
them again.
Open the package carefully and remove the monitor and accessories. Check that the contents are
complete and that the correct options and accessories have been delivered.
If you have any question, please contact your local supplier.
2.2 Mounting the Monitor
Place the monitor on a flat, level surface, hang it on the bed rail, or mount it on a wall. For
detailed information about how to install the wall mount for the monitor, please refer to the
Wall
Mounting Bracket Assembly Instruction
.
WARNING
The wall mounting bracket can be fixed only on a concrete wall.
2.3 Connecting the Power Cable
Connection procedure of the AC power line is listed below:
1.
Make sure the AC power supply complies with the following specifications: 100 V-240 V~,
50 Hz/60 Hz.
2.
Connect the power cord provided with the monitor. Connect the power cord to connector of
the monitor. Connect the other end of the power cord to a grounded power outlet.
NOTE:
1 Connect the power cable to the socket specialized for hospital use.
2 Only use the power cable supplied by EDAN.
2.4 Checking the Monitor
Make sure there is no damage on the measurement accessories and cables. Then turn on the
monitor, check whether the monitor can start normally. Make sure all alarm lamps light up and
the alarm sound is heard when turning on the monitor. Please refer to chapter
Testing Alarms
.