![EDAN iM50 Скачать руководство пользователя страница 53](http://html1.mh-extra.com/html/edan/im50/im50_user-manual_2357173053.webp)
Patient Monitor User Manual Alarms
- 41 -
Chapter 5 Alarms
WARNING
A potential hazard can exist if different alarm presets are used for the same or similar
equipment in any single area, e.g. an intensive care unit or cardiac operating room.
5.1 Alarm Category
The monitor provides two types of alarms: physiological alarms and technical alarms. Also, the
monitor provides prompts.
5.1.1 Physiological Alarms
If one or several physiological parameters of the currently monitored patient exceed the
predefined alarm limit, the monitor will give an alarm, and this type of alarm is called
physiological alarms. About the detailed alarm information, please refer to the Section
Physiological Alarm Information
.
5.1.2 Technical Alarms
If one or several technical status of the device is in abnormal status, the monitor will give an
alarm. And this type of alarm is called technical alarms. Technical alarms can’t be disabled.
About the detailed alarm information, please refer to Section
Technical Alarm Information
.
5.1.3 Prompts
The monitor can give the character indication of monitoring process or other functions. And this
character is called prompts. About the detailed alarm information, please refer to Section
Prompts
.
5.2 Selecting Alarm Tone Type
The user can select the alarm tone type as desired.
1.
Select
Menu
>
Maintenance
>
User Maintain
, and enter the required password.
2.
If configured, select
Alarm Setup
, and set
Alarm Tone
to
Standard
or
Mode 1
.
Standard
: Standard alarm sound according to IEC 60601-1-8.
Mode 1
: User customized alarm sound according to clinical applications.
5.3 Alarm Levels
In terms of severity, the device’s alarm levels can be classified into three categories: high level
alarms, medium level alarms and low level alarms.
1.
High level alarms
A high level alarm intensively warns the operator of a high priority alarm condition which