![EDAN iM50 Скачать руководство пользователя страница 69](http://html1.mh-extra.com/html/edan/im50/im50_user-manual_2357173069.webp)
Patient Monitor User Manual Alarm Information
- 57 -
Message
Cause
Alarm Level
Action Taken
RESP Noise
RR cannot be measured due
to patient movement.
Low
Check whether the
RESP leads are well
connected. Keep the
patient calm for
better monitoring.
RR Exceed
RR measuring value is out
of the measure range.
Medium
Check whether
interference to the
respiratory signal
exists. And check
whether the patient is
breathing normally;
breathing too rapidly
or too slowly may
endanger patient’s
life.
SpO
2
SpO
2
Sensor Off
SpO
2
sensor may be
disconnected from the
patient measuring site.
Low
Make sure the sensor
is well connected to
the patient’s finger
or other parts. Make
sure the monitor and
cables are well
connected.
SpO
2
Comm Fail
SpO
2
module failure or
communication failure
High
Stop using
measuring function
of SpO
2
module, and
notify biomedical
engineer or
manufacturer’s
service staff.
SpO
2
Sensor Err
Malfunction in the SpO
2
sensor or in the extension
cable.
Low
Replace the SpO
2
sensor or the
extension cable.
SpO
2
No Sensor
No SpO
2
sensor was
connected to the monitor.
Low
Make sure the
monitor and sensor
are well connected,
reconnect the sensor.
Содержание iM50
Страница 1: ......
Страница 29: ...Patient Monitor User Manual Basic Operation 17 iM60 iM70 1 2 3 4 5 6 7 8 9 10 11 1 2 3 11 4 5 6 7 8 9 10...
Страница 317: ......