Patient Monitor User Manual Monitoring ECG
- 99 -
When
Electrode Type
is set to
5 Electrodes
or
6 Electrodes
,
Screen Layout
can be set to
Normal
,
Full-Scr
and
Half-Scr
. Select
Normal
to display two ECG waveforms on the main
screen; In 5 electrodes, select
Full-Scr
to display seven ECG waveforms which occupy the area
of seven waveforms on the main screen; In 6 electrodes, select
Full-Scr
to display eight ECG
waveforms which occupy the area of eight waveforms on the main screen; In 5 electrodes, select
Half-Scr
to display seven ECG waveforms on the screen, occupying the area of four waveforms;
In 6 electrodes, select
Half-Scr
to display eight ECG waveforms on the screen, occupying the
area of four waveforms.
When
Electrode Type
is set to
10 Electrodes
,
Screen Layout
can be set to
Normal
and
12
Leads
. Select
Normal
to display two ECG waveforms on the main screen; select
12 Leads
to
display 13 ECG waveforms.
When
Electrode Type
is set to
AUTO
, the monitor can automatically identify the electrode type
according to the actual connection condition of the electrodes, and provide as much lead data as
possible when the condition of the lead signal is satisfied.
NOTE:
1 If
3 Electrodes
is selected in the
ECG Setup
menu, only
Normal
can be selected for
Screen Layout
in the sub-menu.
2 In
10 Electrodes
display interface, the filter can only be set to
Diagnosis
.
3 If
6 Electrodes
is selected in the
ECG Setup
menu, Va and Vb can be respectively
set to either Lead V1 ~ V6, but cannot be set to the same lead, Va is Lead V2 by
default, Vb is Lead V5 by default.
4 If
AUTO
is selected in the
ECG Setup
menu, when the electrodes connected to
patient is reduced from 10 electrodes to 3/5/6 electrodes, user can click
Update
Lead Setup
button to enable the monitor to perform lead off alarm according to
actual electrodes.
5 If
AUTO
is selected in the
ECG Setup
menu, Va and Vb cannot be set when the
monitor recognizes the 10 electrodes system automatically. Va is fixed as V1 and Vb
is fixed as V2.
6
Full-Scr
and
Half-Scr
settings in
Screen Layout
can be stored after restarting. The
user needs to press
shortcut key to switch to the standard screen.
9.6.5 Setting Pace Status
It is important to set the paced status correctly when you start monitoring ECG. To change the
paced status in the ECG Setup menu, select
Pace
to toggle between
On
or
Off
. When
Pace
is set
to
On
:
–
Pace Pulse Rejection is switched on. This means that pacemaker pulses are not counted as
extra QRS complexes.
–
Paced symbol is displayed as
|
on the main screen. At this time, the artifact is displayed on the
screen instead of the actual pacemaker crest. All pacemaker crests are the same, so do not give a
Содержание iM50