Patient Monitor User Manual Alarm Information
- 68 -
Message
Cause
Alarm Level
Action Taken
QuickTemp
QuickTemp Comm
Fail
TEMP module failure or
communication failure.
High
Stop measuring
function of TEMP
module, and notify
biomedical engineer
or manufacturer’s
service staff.
Temp exceed limit
The TEMP value is beyond
the range of +25
°
C ~ +45
°
C.
Medium
Put the sensor into
the sensor bracket,
take it out and
measure again.
No Temp Sensor
TEMP sensor is not
connected to the TEMP
module.
Low
Connect the sensor
and the monitor well,
and measure again.
Ambient temp too
high
The Sensor temperature is
higher than +40
°
C.
Low
Put the sensor into
the sensor bracket,
measure again after
the ambient
temperature reaches
normal value.
Ambient temp too
low
The Sensor temperature is
lower than +10
°
C.
Low
Probe temp too high
The original temperature of
sensor >+33
°
C &
≤
+40
°
C.
Low
Probe data error
Offline: NTC resistance >R
0
°
C; Short: NTC resistance
<R+100
°
C.
Medium
Put the sensor into
the sensor bracket,
take it out and
measure again. If the
problem persists,
stop using measuring
function of TEMP
module, and notify
biomedical engineer
or manufacturer’s
service staff.
Probe heater error
Single failure
Medium
Temp Sensor Off
After the sensor temperature
reaches Predict value, it
descends to the value lower
than Predict value.
Medium
Reconnect the sensor
and make sure that
the cable is properly
connected.