Patient Monitor User Manual Monitoring CO
2
- 149 -
3.
Select
Zero Calibration
in
CO
2
Setup
menu.
4.
After the zeroing calibration is completed, the zeroing message disappears, and the CO
2
monitoring can be performed.
For the Respironics Sidestream CO
2
Module:
1.
Connect the sample line to the module correctly, wait until the monitor’s warm-up message
disappears, keep the inlet of sample line away from CO
2
source.
2.
In the
CO
2
Setup
menu, set
Work Mod
e to
Measure
.
3.
Select
Zero Calibration
in
CO
2
Setup
menu.
4.
After the zeroing calibration is completed, the zeroing message disappears, and the CO
2
monitoring can be performed. If the monitor displays
Breath Detected
or
Zero Required
,
zeroing has failed. Zero calibration must be performed again.
For the Masimo Sidestream CO
2
module:
The highly stable Masimo Sidestream CO
2
module requires no regular zeroing. A room air
reference measurement is performed when the NomoLine is disconnected from the gas inlet,
provided that CO
2
measurements are stable. This zeroing procedure is indicated by the blinking
green.
For the Respironics Mainstream CO
2
Module:
1.
Wait until the monitor’s warm-up message disappears; correctly install the mainstream CO
2
sensor to airway adaptor and remove it from breathing circuit, keep the monitor away from
CO
2
source.
2.
In the
CO
2
Setup
menu, set
Work Mod
e to
Measure
.
3.
Select
Zero Calibration
in
CO
2
Setup
menu.
4.
After the zeroing calibration is completed, the zeroing message disappears, and the CO
2
monitoring can be performed. If the monitor displays
Breath Detected
or
Zero Required
,
zeroing has failed. Zero calibration must be performed again.
Note: CO
2
source includes ventilator, patient’s and operator’s breath.
For Masimo Mainstream CO
2
Module:
The user can click
Zero
button in
CO
2
Setup
,
during the zeroing progress, the monitor displays
prompt information of “
Zeroing
”.
Zeroing needs to be performed only when an offset in gas value is observed, or when an
unspecified accuracy message is displayed.
17.3.2 Sidestream CO
2
Module
17.3.2.1 Measurement Steps
For the EDAN EtCO
2
Module:
1.
Fix the water trap to the water trap holder on the left side of the monitor, confirm it is well