Patient Monitor User Manual Product Specification
- 258 -
Repeatability
0.14
°
C
Note 1: The accuracy consists of two parts, as following:
Accuracy (not including sensor): ± 0.1
°
C (25
°
C ~ 45
°
C) (monitoring mode)
Sensor accuracy:
≤
±
0.2
°
C
NOTE:
The direct mode refers to monitor mode, while adjusted mode refers to predict mode.
A.11 IBP
Complies with IEC 60601-2-34: 2011.
Technique Direct
invasive
measurement
Channel
iM80/M80: 4 channels
iM50/iM60/iM70/M50: 2 channels
IBP
Measure
☆
Measuring
Range
Art
(0 to +300) mmHg
PA/PAWP
(-6 to +120) mmHg
CVP/RAP/LAP/ICP (-10 to +40) mmHg
P1/P2
(-50 to +300) mmHg
Resolution 1
mmHg
☆
Accuracy (not including sensor)
±
2 % or
±
1 mmHg, whichever is greater
ICP:
0 mmHg to 40 mmHg:
±
2 % or
±
1
mmHg, whichever is greater;
-10 mmHg to -1 mmHg: undefined
Pressure Unit
kPa, mmHg, cmH
2
O
Pressure sensor
Sensitivity 5
μ
V/V/mmHg
Impedance Range
300
Ω
to 3000
Ω
Filter
DC~ 12.5 Hz; DC~ 40 Hz
Zero
Range:
±
200 mmHg
Pressure Calibration
Range
IBP (excluding ICP)
80 mmHg to 300 mmHg
ICP
10 mmHg to 40 mmHg
Volume Displacement
7.4 x 10
4
mm
3
/100mmHg