Patient Monitor User Manual Alarm Information
- 75 -
6.4 Adjustable Range of Alarm Limits
ECG alarm limits are listed as follows: unit (bpm)
Patient Type
Adjustable Range
HR
ADU 15~300
PED/NEO 15~350
ST analysis alarm limits are listed as follows: unit (mV)
Adjustable Range
ST -2.0~2.0
QTc and
Δ
QTc alarm limits are listed as follows: unit (ms)
Adjustable Range
QTc 200~800
Δ
QTc 30~200
RESP alarm limits are listed as follows: unit (rpm)
Patient Type
Adjustable Range
RESP
ADU 6~120
PED/NEO 6~150
SpO
2
alarm limits are listed as follows: unit (%)
Adjustable Range
SpO
2
20~100
SpO
2
Desat Limits are listed as follows: unit (%)
Adjustable Range
SpO
2
Desat Limit
20~99
NOTE:
User can set the range through
User Maintain
>
Alarm
Setup
>
SpO
2
Desat Limit
,
SpO
2
Desat Limit should be less than or equal to SpO
2
alarm low limit.
PR
alarm limits are listed as follows: unit (bpm)
Adjustable Range
PR(SpO
2
) EDAN
30~300
Nellcor 30~300
Содержание iM50
Страница 1: ......
Страница 29: ...Patient Monitor User Manual Basic Operation 17 iM60 iM70 1 2 3 4 5 6 7 8 9 10 11 1 2 3 11 4 5 6 7 8 9 10...
Страница 317: ......