Patient Monitor User Manual Default Settings
- 298 -
Unit mmHg
Filter 12.5
Hz
SYS, DIA, MAP
SYS, DIA, MAP
SYS, DIA, MAP
Alarm High Limit (ART, P1, P2)
160, 90, 110
120, 70, 90
90, 60, 70
Alarm Low Limit (ART, P1, P2)
90, 50, 70
70, 40, 50
55, 20, 35
Alarm High Limit (PA)
35, 16, 20
60, 4, 26
60, 4, 26
Alarm Low Limit (PA)
10, 0, 0
24, -4, 12
24, -4, 12
MAP
MAP
MAP
Alarm High Limit (CVP, RAP,
LAP, ICP)
10 4 4
Alarm Low Limit (CVP, RAP,
LAP, ICP)
0 0 0
C.11 CO
2
Default Settings
CO
2
Settings
ADU
PED
NEO
Alarm Switch
On
Alarm Record
Off
Alarm Level
Medium
Work Mode
Standby
Unit mmHg
Apnea Time
20 s
O
2
Compensation
16% (EDAN G2 Module and Respironics Module)
Low (Masimo Module)
N
2
O Compensation
0% (EDAN G2 Module and Respironics Module)
Low (Masimo Module)
Anes Agent
0%
Alarm High Limit (EtCO
2
) 50
50
45
Alarm Low Limit (EtCO
2
) 25
25
30
Alarm High Limit (FiCO
2
) 4
4
4
Alarm High Limit (AwRR)
30
30
100
Alarm Low Limit (AwRR)
8
8
30
Sweep 6.25
mm/s