Patient Monitor User Manual Basic Operation
- 21 -
4
USB interfaces: supports USB 2.0 output. It connects approved USB devices, for
example, USB flash disk, barcode scanner, mouse and keyboard.
5
Multifunctional port: Nurse call port/ analog output/ defibrillator synchronization.
Nurse call port: it connects the monitor to the hospital’s nurse call system. Alarms
indications are alerted through the nurse call system if configured to do so.
Analog output: the monitor outputs the waveform through the port.
Defibrillator synchronization: the monitor outputs the defibrillator
synchronization signal through the port.
6
Network interface. It connects the monitor to MFM-CMS, CMS-LITE or gateway
via standard network cable, which enables MFM-CMS or gateway to achieve
bidirectional communication with the monitor.
7
Anti-theft lock interface
8
Heat sink: Adopt No-fan design, which is dust-free, low noise and low
consumption.
9
Power supply interface: for connecting AC power cable.
10
Power cable safety latch, used to prevent the power cable from loosing or falling.
Place the latch on the power cable and press it down firmly to ensure that it
secures the power cable.
11
Equipotential grounding terminal, if the monitor or other processing unit are used
in internal examinations on the heart, ensure that the room incorporates an
equipotential grounding system to which the monitor and other processing unit
have separate connection.
12 Speaker