Patient Monitor User Manual Monitoring CO
2
- 161 -
2.
Select
Duration
in
Intubation Mode
, there are two selections:
3 min
and
5 min
. The default
setting is
3 min
.
3.
Click
Start
, the monitor will start the intubation mode. During the intubation mode, the
monitor will display the intubation mode and remaining time in the form of text.
When countdown finishes or clicking
End
in
Intubation Mode
menu, the monitor will exit the
intubation mode; After exiting intubation mode, the monitor will respond the physiological alarm
related to CO
2
.
NOTE:
The intubation mode is not available when CO
2
module is in Standby mode.