Patient Monitor User Manual Product Specification
- 251 -
Mode
Manual, Auto, Continuous, Sequence
Measuring Interval in AUTO
Mode (unit: minute)
1/2/3/4/5/10/15/30/60/90/120/180/240/360/480 and User
Define
Continuous
5 min, interval is 5 s
Measuring Parameter
SYS, DIA, MAP, PR
Pressure Unit
kPa, mmHg, cmH
2
O
☆
Measuring Range
☆
Adult Mode
SYS: 25 mmHg to 290 mmHg
DIA: 10 mmHg to 250 mmHg
MAP: 15 mmHg to 260 mmHg
☆
Pediatric Mode
SYS: 25 mmHg to 240 mmHg
DIA: 10 mmHg to 200 mmHg
MAP: 15 mmHg to 215 mmHg
☆
Neonatal Mode
SYS: 25 mmHg to 140 mmHg
DIA: 10 mmHg to 115 mmHg
MAP: 15 mmHg to 125 mmHg
☆
Alarm Type
SYS, DIA, MAP
☆
Cuff Pressure Measuring
Range
0 mmHg to 300 mmHg
Pressure Resolution
1 mmHg
☆
Maximum Mean Error
±5 mmHg
☆
Maximum Standard Deviation 8 mmHg
Maximum Measuring Period
Adult/Pediatric 120
s
Neonate 90
s
Typical Measuring Period
20 s to 35 s (depend on HR/motion disturbance)
Dual Independent Channel Overpressure Protection
Adult (297±3)
mmHg
Pediatric (245±3)
mmHg
Neonatal (147±3)
mmHg
Pre-inflation Pressure